A.9.11. Test Digital Inputs<br />

The next screen is used to check the digital input circuitry. The display shows a<br />

1 if the specific input is closed, 0 if open. If more digital I/O boards are<br />

installed, the NEXT soft key appears, allowing the operator to scroll between<br />

boards. Slots are numbered 1-3; slot 0 is the motherboard.<br />

- TEST SCROLL 3 -<br />

Dig input test<br />

Slot#0 ----00--<br />

NEXT<br />

‘Digit’ (displayed instead of ‘slot#’) identifies the four Digitizer’s inputs. Inputs<br />

are shown from left to right. If a board has 16 inputs, two screens are used to<br />

show the first and the second half, the lower half is shown first.<br />

A.9.12. Test Digital Outputs<br />

This test shows the status of each digital output and allows the operator to force<br />

the output for testing purposes. The output, when forced, stays on until the<br />

CLEAR soft key is pressed or the Run Menu is entered. If an output is forced<br />

and the scroll key is used for reaching some other menu, the output stays in the<br />

forced status until RUN is pressed. This allows the operator to check inputs<br />

while outputs are still in the forced status.<br />

- TEST SCROLL 4 -<br />

Dig output test<br />

output # 1 : ON/OFF<br />

ENTER ON/OFF<br />

Password: Service<br />

To force an output, enter the desired number followed by ENTER. Then use<br />

the ON/OFF key to force it to the ON or OFF status. After the output has been<br />

forced, the CLEAR soft key appears in the middle position.<br />

Slots are numbered 1-3; slot 0 is the motherboard.<br />

WARNING<br />

FORCING THE DIGITAL OUTPUTS MAY CAUSE<br />

MACHINERY TO START. AFTER THE USER<br />

TRIES TO FOURCE AN OUTPUT, THE<br />

FOLLOWING MESSAGE DISPLAYS.<br />

A-46 Weighbridge Physical Parameters <strong>Thermo</strong> Fisher Scientific<br />

REC 4175 Rev B