A.9.11. Test Digital Inputs<br />
The next screen is used to check the digital input circuitry. The display shows a<br />
1 if the specific input is closed, 0 if open. If more digital I/O boards are<br />
installed, the NEXT soft key appears, allowing the operator to scroll between<br />
boards. Slots are numbered 1-3; slot 0 is the motherboard.<br />
- TEST SCROLL 3 -<br />
Dig input test<br />
Slot#0 ----00--<br />
NEXT<br />
‘Digit’ (displayed instead of ‘slot#’) identifies the four Digitizer’s inputs. Inputs<br />
are shown from left to right. If a board has 16 inputs, two screens are used to<br />
show the first and the second half, the lower half is shown first.<br />
A.9.12. Test Digital Outputs<br />
This test shows the status of each digital output and allows the operator to force<br />
the output for testing purposes. The output, when forced, stays on until the<br />
CLEAR soft key is pressed or the Run Menu is entered. If an output is forced<br />
and the scroll key is used for reaching some other menu, the output stays in the<br />
forced status until RUN is pressed. This allows the operator to check inputs<br />
while outputs are still in the forced status.<br />
- TEST SCROLL 4 -<br />
Dig output test<br />
output # 1 : ON/OFF<br />
ENTER ON/OFF<br />
Password: Service<br />
To force an output, enter the desired number followed by ENTER. Then use<br />
the ON/OFF key to force it to the ON or OFF status. After the output has been<br />
forced, the CLEAR soft key appears in the middle position.<br />
Slots are numbered 1-3; slot 0 is the motherboard.<br />
WARNING<br />
FORCING THE DIGITAL OUTPUTS MAY CAUSE<br />
MACHINERY TO START. AFTER THE USER<br />
TRIES TO FOURCE AN OUTPUT, THE<br />
FOLLOWING MESSAGE DISPLAYS.<br />
A-46 Weighbridge Physical Parameters <strong>Thermo</strong> Fisher Scientific<br />
REC 4175 Rev B