Menus<br />

ENTER the number of<br />

belt revolutions to<br />

be timed 000 revs<br />

ENTER ABORT<br />

Default: 1<br />

Min: 1<br />

Max: 100<br />

The next entry is the time per revolution.<br />

Pressing ABORT returns the screen to CAL DATA SCROLL 11.<br />

ENTER the time for<br />

revolutions to pass<br />

reference 000 sec<br />

ENTER ABORT<br />

Default: 30 sec<br />

Min: 1 sec<br />

Max: 16200 sec<br />

When ENTER is pressed, the system times the belt travel according to the<br />

above entered parameters.<br />

Pressing ABORT returns the screen to CAL DATA SCROLL 11.<br />

Timing belt travel<br />

000 sec<br />

ABORT<br />

When test duration test is finished, the new values for length of belt and time are<br />

displayed. Press CONTINUE<br />

A.7.8. Defining Auto Zero Tracking<br />

A periodic auto zero procedure can be automatically executed by the system if<br />

the Auto Zero Tracking option is set to YES. If YES is selected, Auto Zero<br />

Tracking is enabled anytime the belt is running at a flow rate less than the range<br />

setting selected below.<br />

-- CAL DATA SCROLL 12--<br />

Auto zero tracking<br />

> Yes <<br />

ACQ MANUAL<br />

Default: NO<br />

Selections: YES, NO<br />

Password: Operator<br />

A-36 Weighbridge Physical Parameters <strong>Thermo</strong> Fisher Scientific<br />