Menus<br />

When START is pressed, the system automatically moves to the next<br />

instructional screen. The operator presses COUNT each time the mark passes<br />

the reference point until minimum test load conditions are met (refer to the belt<br />

scale installation manual for minimum test level requirements). When the last<br />

revolution passes the mark, the operator must press COUNT followed by<br />

pressing DONE.<br />

Pressing ABORT returns the screen to CAL DATA SCROLL 11.<br />

Press COUNT each<br />

Time a mark passes.<br />

00000 secs 000 revs<br />

COUNT ABORT DONE<br />

ABORT returns the screen to CAL DATA SCROLL 11.<br />

When test duration is finished, the new values for length of belt and time are<br />

displayed. Press CONTINUE<br />

TEST DURATION<br />

Length = 0000.0 Ft<br />

Time = 000 SEC<br />

CONTINUE<br />

A.7.5. Auto Cont Belt Revolutions<br />

If a clip detector is installed and clip detector mode in I/O Definition is set to<br />

"AUTO", the system automatically counts belt revolutions without the need for<br />

pressing the COUNT key. The two screens for Acquiring Test Duration are<br />

displayed as follows:<br />

Start belt. Wait<br />

until 1 st mark<br />

passes reference<br />

ABORT<br />

When the mark passes, the system automatically moves to the next instructional<br />

screen. The system counts each time a mark passes the reference point until the<br />

operator presses the DONE key.<br />

Press DONE when<br />

ready.<br />

0000 secs 000 revs<br />

ABORT DONE<br />

<strong>Thermo</strong> Fisher Scientific Weighbridge Physical Parameters A-33<br />

REC 4175 Rev B