A.7.1.1 Defining the R-Cal Parameters<br />

This section only applies of R-Cal mode was selected as the preferred method.<br />

Enter the resistance in Ohms of the R-cal resistor installed in your Integrator.<br />

Default value installed is 165000 Ohms. See the Data Sheet in front of this<br />

manual for exact resistance.<br />

-- CAL DATA SCROLL 2 --<br />

R-CAL select<br />

Res 165000 ohms<br />

ENTER<br />

Default: 165000 Ohms<br />

Min: 10 Ohms<br />

Max: 1000000 Ohms<br />

The system calculates the CALCON (Calibration Constant) based on the<br />

mechanical and electrical parameters entered in SCALE DATA. This menu is<br />

for reference only.<br />

-- CAL DATA SCROLL 3 --<br />

R-CAL constant<br />

00.00<br />

MENU RUN<br />

A.7.1.2 Defining the Chains Parameters<br />

This section only applies if CHAINS mode was selected as the preferred<br />

method. Enter the weight per foot or meter of the chains used for calibration.<br />

-- CAL DATA SCROLL 4 --<br />

R-CAL constant<br />

00.00<br />

MENU RUN<br />

Password:<br />

Service<br />

Password:<br />

Service<br />

English/Mixed Metric<br />

Default: 0.000 Lbs/Ft Default: 0.000 kg/m<br />

Min: 0.000 Lbs/Ft Min: 0.000 kg/m<br />

Max: 1000.000 Lbs/FT Max: 3000 kg/m<br />

The system calculates the CALCON based on the mechanical and electrical<br />

parameters entered in SCALE DATA. This menu is for reference only.<br />

-- CAL DATA SCROLL 5 --<br />

CHAIN CAL constant<br />

00.00<br />

MENU RUN<br />