Menus<br />

A.6.8. Setting Dead Band<br />

The dead band is a percentage of the scale capacity (rate) in which the rate is<br />

ignored (if any) and a zero rate is forced. In addition, Totalization is frozen<br />

when rate is below dead band.<br />

Default: 0%<br />

Min: 0%<br />

Max: 5%<br />

A.6.9. Weights and Measures Mode (Only if Audit Trail Option is<br />

Installed)<br />

This scroll details certain parameters to conform to the approval agency.<br />

Default: NONE<br />

Choices: NONE, OIML<br />

A.7. Calibration Data -<br />

-- SC DATA SCROLL 10 --<br />

Zero dead-Band Range<br />

0 %<br />

ENTER<br />

-- SC DATA SCROLL 11 --<br />

W&M mode<br />

>NONE<<br />

CHOICE ENTER<br />

The CAL DATA scroll allows the operator to enter or set parameters, which<br />

relate to the calibration of the scale.<br />

A.7.1. Defining the Calibration Mode<br />

Select the method of simulated automatic calibration. The selected method is<br />

the only one displayed in the calibration section of MENU 1.<br />

-- CAL DATA SCROLL 1 --<br />

Calibration mode:<br />

< R-CAL<<br />

CHOICE ENTER<br />

Default: R-CAL<br />

Choice: R-CAL, CHAIN, WEIGHTS<br />

Password:<br />

Service<br />

Password:<br />

Service<br />

A-28 Weighbridge Physical Parameters <strong>Thermo</strong> Fisher Scientific<br />

REC 4175 Rev B