A.3.3.4 Executing the Span Calibration<br />
Whichever method has been used to start automatic span calibration, after start<br />
is pressed, the following screen displays:<br />
AUTO SPANNING<br />
Time remaining 0000<br />
Rate 0000.0 Tph<br />
Total 000000 Tons<br />
During Auto Span, the resolution of the total is ten times higher than normal.<br />
The time remaining shown in Line 2 is calculated upon the current speed and is<br />
based on the test duration.<br />
If the belt is stopped during the test, a message is displayed indicating the<br />
procedure has been suspended.<br />
No action is required from the operator; wait until the test is completed.<br />
A.3.3.5 Material Factor<br />
Use this step only if a material calibration is done before and the current<br />
simulated load method has no material factor installed.<br />
Note: This section does not apply if one of the following exists:<br />
1. A calibration with material has not been run before,<br />
2. A manual span entry is done, or<br />
3. This is not the first time the current simulated method is used.<br />
Proceed to Section 3.10.3.1.6.<br />
It is very important to understand when this procedure is executed; the system<br />
does not alter the span. The span is assumed correct because it was obtained<br />
from a test with material.<br />
The system acquires the Material Factor for the current calibration method<br />
instead. This means the Integrator knows in the future how to use this method<br />
for correctly changing the system’s span.<br />
Use the following steps to incorporate the Material Factor. The following<br />
screen displays:<br />
AUTO SPAN COMPLETE<br />
Error +/-00.00 %<br />
Unfactored Calcon<br />
EXIT FACTOR REPEAT<br />