6. Restart the conveyor belt if it is not running<br />

7. Press START<br />

8. Continue to Section 3.10.3.1.4<br />

A.3.3.4 Executing the Span Calibration<br />

Whichever method has been used to start automatic span calibration, after start<br />

is pressed, the following screen displays:<br />

AUTO SPANNING<br />

Time remaining 0000<br />

Rate 0000.0 Tph<br />

Total 000000 Tons<br />

During Auto Span, the resolution of the total is ten times higher than normal.<br />

The time remaining shown in Line 2 is calculated upon the current speed and is<br />

based on the test duration.<br />

If the belt is stopped during the test, a message is displayed indicating the<br />

procedure has been suspended.<br />

No action is required from the operator; wait until the test is completed.<br />

A.3.3.5 Material Factor<br />

Use this step only if a material calibration is done before and the current<br />

simulated load method has no material factor installed.<br />

Note: This section does not apply if one of the following exists:<br />

1. A calibration with material has not been run before,<br />

2. A manual span entry is done, or<br />

3. This is not the first time the current simulated method is used.<br />

Proceed to Section 3.10.3.1.6.<br />

It is very important to understand when this procedure is executed; the system<br />

does not alter the span. The span is assumed correct because it was obtained<br />

from a test with material.<br />

The system acquires the Material Factor for the current calibration method<br />

instead. This means the Integrator knows in the future how to use this method<br />

for correctly changing the system’s span.<br />

Use the following steps to incorporate the Material Factor. The following<br />

screen displays:<br />

AUTO SPAN COMPLETE<br />

Error +/-00.00 %<br />

Unfactored Calcon<br />

EXIT FACTOR REPEAT<br />

A-8 Weighbridge Physical Parameters <strong>Thermo</strong> Fisher Scientific<br />

REC 4175 Rev B