Maintenance<br />
4-10<br />
8. Press ENTER twice. The display should respond with NEW PASSWORD<br />
ACQUIRED.<br />
9. Scroll down to OPERATOR PASSWORD. Press ENTER twice. The<br />
display should respond with NEW PASSWORD ACQUIRED.<br />
10. Select MAIN MENU 3. The PROT soft key should not appear, indicating<br />
all passwords have been erased. If PROT does appear, repeat Steps 1<br />
through 9.<br />
11. See Section for entering new passwords.<br />
4.9 Lithium Battery Replacement<br />
The Micro-Tech volatile memory backup battery can be replaced without any<br />
special tools.<br />
WARNING<br />
DANGER OF EXPLOSION IF BATTERY IS<br />
INCORRECTLY REPLACED.<br />
Replace only with same or equivalent type recommended by <strong>Thermo</strong> Scientific.<br />
Dispose of used battery according to manufactures instruction on battery or<br />
return to <strong>Thermo</strong> Scientific. (Refer to Section 4.10.)<br />
1. Record all configuration, setup, and calibration data before removing<br />
battery. All information is lost when the battery is removed.<br />
2. Turn the Micro-Tech power off at the mains.<br />
3. Remove the battery from its compression socket.<br />
4. Observe the polarity markings on the battery socket base before inserting the<br />
new battery. The lithium battery is .3V, 1.2 AH, 2/3 A, <strong>Thermo</strong> Scientific<br />
part number 037188.<br />
5. Insert battery<br />
6. Restore power to the Micro-Tech.<br />
7. Cold start the Micro-Tech. See Section 4.4 for cold start procedures.<br />
8. Re-enter all data recorded in Step 1.<br />
4.10 Disposal of Hazardous Waste<br />
Disposal of Lithium batteries and soldered print circuit boards should be in<br />
accordance with your local Hazardous Waste Policy.<br />
As an alternative, you may return product supplied by <strong>Thermo</strong> Scientific, freight<br />
prepaid for disposal. Contact <strong>Thermo</strong> Scientific Repair Department for a Return<br />
Authorization Number before shipping any product for disposal.<br />
Replacement Parts <strong>Thermo</strong> Fisher Scientific<br />
REC 4175 Rev B