Maintenance<br />

4-8<br />

Press START, the following screens display in sequence:<br />

The message “Test PASSED” is displayed if the test runs correctly. If<br />

something wrong is detected, the message “Test FAILED” is displayed and the<br />

soft key CONTINUE is shown. Press CONTINUE and move to the next test.<br />

If the internal test has failed, call <strong>Thermo</strong> Scientific Customer Service.<br />

4.6 Load Cell Excitation and Signal Voltage<br />

1. Measure excitation voltage across terminal 21 negative and 20 positive in<br />

the scale junction box. This should be 10 VDC ± 5%.<br />

2. If the excitation voltage is incorrect then measure the excitation voltage in<br />

the Integrator across terminal TB4-33 negative and the TB4-32 positive.<br />

This should be 10 VDC ± 5%.<br />

3. Measure DC millivolt signal voltage across terminal 22 positive and 23<br />

negative in the scale junction box. This should be within 0-30 millivolts DC<br />

(3 mV/V load cell).<br />

4. Measure DC millivolt signal voltage across terminal TB3-30 positive and<br />

TB4-31 negative in the Integrator. This should be the same as Step 3<br />

above.<br />

5. The millivolt output is in direct relation to weight applied. As weight is<br />

increased, output should increase.<br />

4.7 Resetting Master Total<br />

-- TEST SCROLL 2A --<br />

Testing ROM<br />

Test PASSED<br />

-- TEST SCROLL 2B --<br />

Testing RAM<br />

Test PASSED<br />

-- TEST SCROLL 2C --<br />

Testing E2PROM<br />

Test Passed<br />

Displays only if Audit Trail<br />

is installed.<br />

Use the following steps to reset the Master Total or the Remote Counter<br />

Overflow.<br />

Replacement Parts <strong>Thermo</strong> Fisher Scientific<br />

REC 4175 Rev B