25.04.2013 Views

Thermo Ramsey - Hoferick Engineering GmbH

Thermo Ramsey - Hoferick Engineering GmbH

Thermo Ramsey - Hoferick Engineering GmbH

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Maintenance<br />

4-6<br />

4.4.1 Forcing a Cold Start from Run Mode<br />

Use the following steps to force a cold start from RUN mode.<br />

1. Press and hold at the same time the LEFT HAND ARROW and the CLEAR<br />

keys until the following screen displays.<br />

Install Factory<br />

Defaults?<br />

NO YES<br />

No, Returns to Run Mode<br />

Yes, the following screen displays:<br />

-- MEMORY ERASED --<br />

Choose the language<br />

key to continue to<br />

ESP USA<br />

2. When this screen displays, all field entry data has been replaced by the<br />

factory default constants. Proceed to Section 2.5and follow the Initial Setup<br />

procedures.<br />

Note: If the software corruption was catastrophic and the memory will not erase<br />

do the following:<br />

Press and hold in the LEFT ARROW and the CLEAR key. While holding in<br />

both keys, cycle line power. In the event the MEMORY ERASED screen does<br />

not appear, consult the factory.<br />

4.4.2 Forcing a Cold Start from the Diagnostic Scroll<br />

Use the following steps to force a cold start from the DIAGNOSITICS<br />

SCROLL:<br />

1. Press MENU until MAIN MENU 3 appears<br />

2. Press the DIAG soft key<br />

3. Press the DOWN ARROW until the following screen is displayed:<br />

-- DIAGNOST.SCROLL 15 --<br />

Force cold start<br />

ENTER<br />

Replacement Parts <strong>Thermo</strong> Fisher Scientific<br />

REC 4175 Rev B