Maintenance<br />
4-4<br />
14. Calib Time<br />
If a calibration check time is entered and the time expires, this alarm occurs.<br />
The purpose is to remind the operator that the calibration has not been<br />
checked for a considerably long period.<br />
Check Calibration<br />
15. Ext. Alarm 1<br />
Digital inputs can be programmed to detect external alarm conditions such<br />
as emergency switches, max level switches or other. This alarm is<br />
associated to the external alarm #1.<br />
Check External alarm #1.<br />
16. Ext. Alarm 2<br />
Check external alarm #2.<br />
17. Ext. Alarm 3<br />
Check external alarm #3<br />
18. Overflow Tot.<br />
This message indicates the output pulse generator for the remote mechanical<br />
totalizer has reached an overflow condition. The rate may be too high or the<br />
pulse divider has been set to small.<br />
Check the rate.<br />
Check and eventually increase the pulse divider.<br />
This message is also displayed if the Master Total rolls over.<br />
19. AZT Limit<br />
The Auto Zero Tracking function has reached the maximum limit of tare<br />
that is allowed to be autmatically cleared. The scale may be dirty or<br />
misaligned.<br />
Check and clean the scale.<br />
Perform an Auto Zero.<br />
20. Batch Deviat.<br />
This alarm is only visible if the optional Load Out board is installed. It is<br />
generated when a batch is run and the error detected at the end of the batch<br />
is higher than specified.<br />
Check the parameters of the load out.<br />
Adjust the preset value.<br />
21. BCD Error<br />
This message is only displayed if the optional BCD output board is installed.<br />
If the variable to be converted in the BCD format has more tha 4 digits, the<br />
alarm is generated.<br />
Check the size of variables and the BCD data setup.<br />
Replacement Parts <strong>Thermo</strong> Fisher Scientific<br />
REC 4175 Rev B