Maintenance<br />

4-4<br />

14. Calib Time<br />

If a calibration check time is entered and the time expires, this alarm occurs.<br />

The purpose is to remind the operator that the calibration has not been<br />

checked for a considerably long period.<br />

Check Calibration<br />

15. Ext. Alarm 1<br />

Digital inputs can be programmed to detect external alarm conditions such<br />

as emergency switches, max level switches or other. This alarm is<br />

associated to the external alarm #1.<br />

Check External alarm #1.<br />

16. Ext. Alarm 2<br />

Check external alarm #2.<br />

17. Ext. Alarm 3<br />

Check external alarm #3<br />

18. Overflow Tot.<br />

This message indicates the output pulse generator for the remote mechanical<br />

totalizer has reached an overflow condition. The rate may be too high or the<br />

pulse divider has been set to small.<br />

Check the rate.<br />

Check and eventually increase the pulse divider.<br />

This message is also displayed if the Master Total rolls over.<br />

19. AZT Limit<br />

The Auto Zero Tracking function has reached the maximum limit of tare<br />

that is allowed to be autmatically cleared. The scale may be dirty or<br />

misaligned.<br />

Check and clean the scale.<br />

Perform an Auto Zero.<br />

20. Batch Deviat.<br />

This alarm is only visible if the optional Load Out board is installed. It is<br />

generated when a batch is run and the error detected at the end of the batch<br />

is higher than specified.<br />

Check the parameters of the load out.<br />

Adjust the preset value.<br />

21. BCD Error<br />

This message is only displayed if the optional BCD output board is installed.<br />

If the variable to be converted in the BCD format has more tha 4 digits, the<br />

alarm is generated.<br />

Check the size of variables and the BCD data setup.<br />

Replacement Parts <strong>Thermo</strong> Fisher Scientific<br />

REC 4175 Rev B