Maintenance 4.3 Alarm Messages 4-2 The ALARM message is assigned to the right hand soft key when an alarm is pending. The Alarm message and its LED flash at the same time. The following screen is displayed when the right hand soft key is pressed. ALARM NEW xxxxxxxxxxxxxxxxxxxxxx MM-DD-YYYY HH:MM RESET NEXT • NEW indicates an alarm that has not yet been acknowledged. When the operator presses RESET to clear the alarm, the alarm disappears only if the trigger for the alarm does not exist any longer. If the alarm is still pending, ACK is displayed instead of NEW. • NEXT is used to scroll between the pending alarms. • XXXXXXXXXXXXXXXXX represents one of the conditions listed in Section.4.3.1. 4.3.1 Alarms List 1. Clock Fail The system has detected a failure on the clock calendar circuit. Go to the DIAGNOSTICS screen and re-enter the date and time. Check the battery Replace the motherboard. 2. Load Cell Fail The system has detected an error on the load cell signal. Check the load cell connections. Check the load cell(s). 3. RAM Fail The system has detected an error on the RAM (Random Access Memory) checksum during the internal periodic test. The RAM is used to store variables and set up data. Replace the motherboard. 4. ROM Fail The system has detected a failure on the ROM (Read Only Memory) checksum during the internal periodic test. The ROM is used to store the program. Replace the mother board Replacement Parts Thermo Fisher Scientific REC 4175 Rev B

Replacement Parts 5. High Load The belt load has been detected to be higher than the maximum belt loading entered in the ALARM SET UP menu. 6. Low Load The belt load has been detected to be lower than the minimum belt loading entered in the ALARM SET UP men. 7. High Rate The rate has been detected to be higher than the maximum rate entered in the ALARM SET UP menu. 8. Low Rate The rate has been detected to be lower than the minimum rate entered in the ALARM SET UP menu. 9. High Speed The speed has been detected to be higher than the maximum rate entered in the ALARM SET UP menu. 10. Low Speed The speed has been detected to be lower than the minimum speed entered in the ALARM SET UP menu. 11. Warm Start The system has detected a power loss condition, or power was removed for an undefined period. 12. Cold Start The system has detected the loss of the set up data after power was removed. The instrument needs to be setup and calibrated. Replace the motherboard Note: The message COLD START never appears on the screen. This is because a Cold Start forces a start up procedure to be executed, and the alarm itself is cleared after the se up is completed. However, the alarm LED and the digital output will be showing an alarm during the initial set up procedure. 13. P.D. Calibrate When the system is powered off while a calibration sequence is in progress, the scale may not be properly calibrated. Check calibration Thermo Fisher Scientific Replacement Parts REC 4175 Rev B 4-3

Maintenance<br />

4.3 Alarm Messages<br />

4-2<br />

The ALARM message is assigned to the right hand soft key when an alarm is<br />

pending. The Alarm message and its LED flash at the same time.<br />

The following screen is displayed when the right hand soft key is pressed.<br />

ALARM NEW<br />

xxxxxxxxxxxxxxxxxxxxxx<br />

MM-DD-YYYY HH:MM<br />

RESET NEXT<br />

• NEW indicates an alarm that has not yet been acknowledged. When the<br />

operator presses RESET to clear the alarm, the alarm disappears only if the<br />

trigger for the alarm does not exist any longer. If the alarm is still pending,<br />

ACK is displayed instead of NEW.<br />

• NEXT is used to scroll between the pending alarms.<br />

• XXXXXXXXXXXXXXXXX represents one of the conditions listed in<br />

Section.4.3.1.<br />

4.3.1 Alarms List<br />

1. Clock Fail<br />

The system has detected a failure on the clock calendar circuit.<br />

Go to the DIAGNOSTICS screen and re-enter the date and time.<br />

Check the battery<br />

Replace the motherboard.<br />

2. Load Cell Fail<br />

The system has detected an error on the load cell signal.<br />

Check the load cell connections.<br />

Check the load cell(s).<br />

3. RAM Fail<br />

The system has detected an error on the RAM (Random Access Memory)<br />

checksum during the internal periodic test. The RAM is used to store<br />

variables and set up data.<br />

Replace the motherboard.<br />

4. ROM Fail<br />

The system has detected a failure on the ROM (Read Only Memory)<br />

checksum during the internal periodic test. The ROM is used to store the<br />

program.<br />

Replace the mother board<br />