PROCEDURE<br />

1. Press and hold ON button until colorimeter turns on.<br />

2. Press ENTER to start.<br />

3. Press ENTER to select <strong>TEST</strong>ING MENU.<br />

4. Select ALL <strong>TEST</strong>S (or another sequence containing 39 DO)from<br />

<strong>TEST</strong>ING MENU.<br />

5. Scroll to and select 39 DO from menu.<br />

6. Rinse a clean tube (0290) with untreated sample water. Fill to the 10 mL<br />

line with sample. This tube is the BLANK.<br />

7. Insert tube into chamber, close lid and select SCAN BLANK.<br />

8. Fill a second tube (0290) to the 10 line with the treated “Fixed” sample.<br />

This tube is the SAMPLE.<br />

9. Remove BLANK from colorimeter, insert SAMPLE tube into chamber,<br />

close lid and select SCAN SAMPLE. Record result.<br />

10. Press OFF button to turn colorimeter off or press EXIT buttontoexittoa<br />

previous menu or make another menu selection.<br />

Smart2 <strong>TEST</strong> PROCEDURES 2.04 Dissolved Oxygen 3/3