ALUMINUM<br />

ERIOCHROME CYANINE R METHOD CODE 364I-SC<br />

QUANTITY CONTENTS CODE<br />

5 g *Aluminum Inhibitor Reagent *7865-C<br />

2 x 120 mL *Aluminum Buffer Reagent *7866-J<br />

120 mL Aluminum Indicator Reagent 7867-J<br />

15 mL Aluminum Complexing Reagent 7868-E<br />

1 Spoon, 0.05 g, plastic 0696<br />

2 Pipets, 1.0 mL, plastic 0354<br />

1 Test Tube, glass, 5 mL w/cap 0230<br />

*WARNING: Reagents marked with a * are considered to be potential health hazards.<br />

To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS<br />

CD or our web site. To obtain a printed copy, contact us by e-mail, phone or fax.<br />

Aluminum is the third most common element in the earth’s crust, which<br />

accounts for its wide appearance in many water supplies. Aluminum exists in<br />

water as soluble salts, colloidal compounds, and insoluble compounds. In<br />

wastewater that has been treated by alum coagulation it will appear in one or<br />

more of the above forms. Properly treated drinking water should have an<br />

aluminum concentration below 0.05 mg/L.<br />

APPLICATION: Drinking, surface, and saline waters; domestic and industrial<br />

wastewater.<br />

RANGE: 0.00–0.30 ppm Aluminum<br />

METHOD: Aluminum ions buffered to a pH of 6.0 react with<br />

Eriochrome Cyanine R dye to produce a pink to red complex<br />

SAMPLE<br />

HANDLING &<br />

PRESERVATION:<br />

in proportion to the concentration.<br />

Collect sample in acid washed glass or plastic bottle. Analyze<br />

as soon as possible.<br />

INTERFERENCES: Fluoride and polyphosphate will interfere. Interference from<br />

iron and manganese is eliminated by the addition of an<br />

inhibitor.<br />