PROCEDURE<br />

1. Press and hold ON burton until colorimeter turns on.<br />

2. Press ENTER to start.<br />

3. Press ENTER to select <strong>TEST</strong>ING MENU.<br />

4. Select ALL <strong>TEST</strong>S (or another sequence containing 28 Color)from<br />

<strong>TEST</strong>ING MENU.<br />

5. Scroll to and select 28 Color from menu.<br />

6. Rinse a tube (0290) with color-free water (distilled or deionized water).<br />

Fill to 10 mL line with color-free water.<br />

7. Insert the tube into chamber, close lid and select SCAN BLANK.<br />

8. Remove tube from colorimeter. Empty tube.<br />

9. Rinse tube with sample water. Fill to 10 mL line with water sample.<br />

10. Insert tube with sample water, close lid and select SCAN SAMPLE. Record<br />

result in color units.<br />

11. Press OFF button to turn the colorimeter off or press EXIT button to exit<br />

to a previous menu or make another menu selection.<br />

Color 2/2 Smart2 <strong>TEST</strong> PROCEDURES 2.04