PROCEDURE<br />

Use COD adapter (see p. 22).<br />

1. Homogenize sample if necessary.<br />

2. Preheat COD heater block to 150±2°C.<br />

3. Remove cap from COD tube vial. Hold vial at a 45° angle. Use a<br />

volumetric pipet, to carefully add 2.0 mL sample water allowing the<br />

sample to run down the side of the vial.<br />

4. Cap and mix thoroughly.<br />

5. Rinse the outside of the vial with distilled water. Wipe dry with a paper<br />

towel.<br />

6. Repeat steps 3 through 5 using 2.0 mL distilled water. This is the reagent<br />

blank.<br />

7. Place vials in preheated COD block heater and maintain temperature at<br />

150±2°C for two hours.<br />

8. At the end of the heating period turn the heater off. Wait 20 minutes for<br />

the vials to cool to 120°C or less.<br />

9. Remove vials from block heater. Invert several times to mix.<br />

10. Allow to cool to room temperature.<br />

11. Press and hold ON button until colorimeter turns on.<br />

12. Press ENTER to start.<br />

13. Press ENTER to select <strong>TEST</strong>ING MENU.<br />

14. Select ALL <strong>TEST</strong>S (or another sequence containing 25 COD LR)from<br />

PROGRAMMED <strong>TEST</strong>S menu.<br />

15. Scroll to and select 25 COD LR from menu.<br />

16. Wipe the blank vial with a damp towel to remove fingerprints and<br />

smudges. Wipe with a dry towel.<br />

17. Insert reagent blank tube into chamber. Select SCAN BLANK.<br />

18. Remove tube from colorimeter.<br />

19. Insert digested water sample tube into chamber. Select SCAN SAMPLE.<br />

Record result. For the most accurate results, take three readings on each<br />

sample and average the results.<br />

20. Press OFF button to turn colorimeter off or press EXIT button to exit to<br />

a previous menu or make another menu selection.<br />

Smart2 <strong>TEST</strong> PROCEDURES 03.05 COD-LR 3/4