COD–LOW RANGE<br />
MERCURY FREE DIGESTION CODE 0072-SC<br />
MERCURY DIGESTION CODE 0075-SC<br />
QUANTITY CONTENTS CODE<br />
25 *COD Low Range Mercury Free Tubes *0072-SC<br />
or 25 *COD Low Range Mercury Tubes *0075-SC<br />
*WARNING: Reagents marked with an * are considered hazardous substances. To<br />
view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD<br />
or our web site. To obtain a printed copy, contact us by e-mail, phone or fax.<br />
COD Low Range Mercury Free Tubes are not USEPA approved.<br />
COD Low Range Mercury Tubes are USEPA approved.<br />
<strong>Equipment</strong> needed but not supplied:<br />
QUANTITY CONTENTS CODE<br />
1 COD Adapter 5-0087<br />
1 COD Reactor, 12 vial, 110V 5-0102<br />
or 1 COD Reactor, 12 vial, 230V 5-0102-EX2<br />
1 Measuring Pipet, 1.0 mL 2-2110<br />
1 Pipet Bulb 2-2164<br />
Chemical Oxygen Demand (COD) is a measure of the amount of organic<br />
matter in water which is susceptible to oxidation by chemical oxidants. COD<br />
can be empirically related to the Biological Oxygen Demand (BOD) and<br />
organic carbon content of a specific source of water. This correlation must be<br />
determined experimentally for each source of water.<br />
APPLICATION: Domestic and industrial wastes.<br />
RANGE: 5–150 mg/L COD<br />
METHOD: Dichromate in the presence of silver salts, at high<br />
temperature in a closed system, oxidizes most organic<br />
compounds to 95-100% of the theoretical amount. This<br />
process is called digestion. As dichromate oxidizes the<br />
organic compounds, the amount of yellow color is reduced.<br />
The remaining yellow color is measured colorimetrically at<br />
the 420 nm and is directly proportional to the COD of the<br />
SAMPLE<br />
HANDLING &<br />
PRESERVATION:<br />
sample.<br />
Collect samples in glass and test as soon as possible. If<br />
samples must be stored, preservation is accomplished by the<br />
addition of concentrated H2SO4 to adjust the pH below 2.<br />
Samples with suspended solids should be homogenized in a<br />
blender (100 mL for 30 seconds) and then stirred gently<br />
with a magnetic stirrer.<br />
Smart2 <strong>TEST</strong> PROCEDURES 03.05 COD–LR 1/4