24.04.2013 Views

TEST INSTRUCTIONS - Geotech Environmental Equipment

TEST INSTRUCTIONS - Geotech Environmental Equipment

TEST INSTRUCTIONS - Geotech Environmental Equipment

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

PROCEDURE<br />

1. Press and hold ON button until colorimeter turns on.<br />

2. Press ENTER to start.<br />

3. Press ENTER to select <strong>TEST</strong>ING MENU.<br />

4. Select ALL <strong>TEST</strong>S (or another sequence containing 23 Chromium-TT)<br />

from <strong>TEST</strong>ING MENU.<br />

5. Scroll to and select 23 Chromium-TT from menu.<br />

6. Rinse a tube (0290) with sample water. Fill to 10 mL line with sample.<br />

7. Insert the tube into chamber, close lid and select SCAN BLANK.<br />

8. Remove the tube from colorimeter.<br />

9. Add one *Chromium Spectrophotometric Grade Tablet (3889A).<br />

10. Use Tablet Crusher (0175) to crush tablet.<br />

11. Cap tube.<br />

12. Shake vigorously for 30 seconds.<br />

13. Wait 3 minutes.<br />

14. Insert tube into chamber, close lid and select SCAN SAMPLE. Record<br />

result in ppm chromium.<br />

15. Press OFF button to turn the colorimeter off or press EXIT button to exit<br />

to a previous menu or make another menu selection.<br />

NOTES: For best possible results, a reagent blank should be determined to<br />

account for any contribution to the test result by the reagent system. To<br />

determine the reagent blank, follow the above test procedure to scan a distilled<br />

or deionized water blank. Then follow the above procedure to perform the test<br />

on a distilled or deionized water sample. This test result is the reagent blank.<br />

Subtract the reagent blank from all subsequent test results of unknown samples.<br />

–2<br />

To convert results to ppm chromate (CrO ), multiply by 2.23. To convert<br />

4<br />

result to ppm sodium chromate (Na CrO ) multiply by 3.12.<br />

2 4<br />

Chromium-TesTab 1/2 Smart2 <strong>TEST</strong> PROCEDURES 1.07