TEST INSTRUCTIONS - Geotech Environmental Equipment
TEST INSTRUCTIONS - Geotech Environmental Equipment
TEST INSTRUCTIONS - Geotech Environmental Equipment
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
PROCEDURE<br />
Use 10 mm square cell adapter.<br />
1. Press and hold ON button until colorimeter turns on.<br />
2. Press ENTER to start.<br />
3. Press ENTER to select Testing Menu.<br />
4. Select ALL <strong>TEST</strong>S (or another sequence containing 16 Cl Free-UDV)<br />
from <strong>TEST</strong>ING MENU.<br />
5. Scroll to and select 16 Cl Free-UDV from menu.<br />
6. Rinse a clean vial (0156) with sample water.<br />
7. Use the syringe (1184) to add 3mL of sample to the vial.<br />
8. Insert the vial into chamber, close the lid and select SCAN BLANK.<br />
9. Remove the vial from the colorimeter.<br />
10. Use the syringe (1184) to add 3mL of sample to a *Free Chlorine UDV<br />
vial (4311).<br />
11. Shake vigorously until powder dissolves completely.<br />
NOTE: If powder residue remains in the bottom of the vial after inverting<br />
or air bubbles form, invert once more and tap bottom of vial sharply once<br />
or twice to dislodge powder and bubbles. Mix.<br />
12. Immediately insert tube into chamber, close lid and select SCAN SAMPLE.<br />
Record result in ppm free chlorine.<br />
13. Press OFF button to turn the colorimeter off or press EXIT button to exit<br />
to a previous menu or make another menu selection.<br />
NOTES: For best possible results, a reagent blank should be determined to<br />
account for any contribution to the test result by the reagent system. To<br />
determine the reagent blank, follow the above test procedure to scan a distilled<br />
or deionized water blank. Then follow the above procedure to perform the test<br />
on a distilled or deionized water sample. This test result is the reagent blank.<br />
Subtract the reagent blank from all subsequent test results of unknown samples.<br />
It is necessary to determine the reagent blank only when a new lot number of<br />
reagents are obtained.<br />
UDVs from opened pouches should be used promptly. Store unused vials from<br />
opened pouches in the Foil Storage Bag (9467) to extend the shelf life of the<br />
reagent. Generally, UDVs stored in the bag should be used within 10 days if<br />
the humidity is less than 50% and within 5 days if humidity is greater than<br />
50%. The Foil Storage Bag contains a desiccant pack with indicator. When<br />
the indicator in the window turns from blue to pink, the bag should be<br />
replaced.<br />
Smart2 <strong>TEST</strong> PROCEDURES 2.04 Chlorine, Free–UDV 3/3