PROCEDURE–FREE CHLORINE<br />

1. Press and hold ON button until colorimeter turns on.<br />

2. Press ENTER to start.<br />

3. Press ENTER to select <strong>TEST</strong>ING MENU.<br />

4. Select ALL <strong>TEST</strong>S (or another sequence containing 15 Chlorine)<br />

from <strong>TEST</strong>ING MENU.<br />

5. Scroll to and select 15 Chlorine from menu.<br />

6. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with<br />

sample.<br />

7. Insert tube into chamber, close lid and select SCAN BLANK.<br />

8. Remove tube from colorimeter and pour off all but a sufficient amount of<br />

sample water to cover a tablet. Add one *Chlorine DPD #1 Instrument<br />

Grade Tablet (6903). Crush tablet with a tablet crusher (0175), then add<br />

sample water until tube is filled to 10 mL line. Cap tube and shake until<br />

tablet has dissolved. Solution will turn pink if free chlorine is present.<br />

Wait 15 seconds, but no longer than 30 seconds. Mix.<br />

9. Insert tube into chamber, close lid and select SCAN SAMPLE.<br />

PROCEDURE–COMBINED CHLORINE<br />

10. Add one *Chlorine DPD #3 Instrument Grade Tablet (6197) to sample<br />

from Step 8 above. Crush tablet with tablet crusher (0175). Cap tube and<br />

shake until tablet dissolves. An increase in color represents combined<br />

chlorine.<br />

NOTE: For wastewater samples, Standard Methods for the Examination of<br />

Water and Wastewater recommends waiting 2 minutes for full color<br />

development.<br />

11. Insert sample into chamber, close lid and select SCAN SAMPLE. Record<br />

result as Total Chlorine.<br />

12. Subtract free chlorine reading from total chlorine reading to obtain<br />

concentration of combined chlorine.<br />

13. Press the OFF button to turn off the colorimeter or press the EXIT button<br />

to exit to a previous menu or make another menu selection.<br />