SAMPLE HANDLING & PRESERVATION: Chlorine in aqueous solutions is not stable, and the chlorine content of samples or solutions, particularly weak solutions, will rapidly decrease. Exposure to sunlight or agitation will accelerate the reduction of chlorine present in such solutions. For best results, start analysis immediately after sampling. Samples to be analyzed for chlorine cannot be preserved or stored. INTERFERENCE: The only interfering substance likely to be encountered in water is oxidized manganese. The extent of this interference can be determined by treating a sample with sodium arsenite to destroy the chlorine present so that the degree of interference can be measured. Iodine and bromine can give a positive interference, but these are not normally present unless they have been added as sanitizers. Chlorine- Liquid DPD 2/3 Smart2 TEST PROCEDURES 2.04
PROCEDURE–FREE CHLORINE 1. Press and hold ON button until colorimeter turns on. 2. Press ENTER to start. 3. Press ENTER to select Testing Menu. 4. Select ALL TESTS (or another sequence containing 17 Cl DPD-Liq) from TESTING MENU. 5. Scroll to and select 17 Cl DPD-Liq from menu. 6. Rinse a tube (0290) with sample water. Fill to 10 mL with sample. 7. Insert the tube into chamber, close lid and select SCAN BLANK. 8. Remove the tube from colorimeter. 9. Add 5 drops of DPD 1A Free Chlorine Reagent (P-6740). 10. Add 5 drops of *DPD 1B Free Chlorine Reagent (P-6741). Cap and mix. 11. Insert tube into chamber, close lid and select SCAN SAMPLE. Record result as ppm free chlorine. PROCEDURE–TOTAL CHLORINE 12. Add 5 drops of *DPD 3 Total Chlorine Reagent (P-6741). Cap and mix. NOTE: For wastewater samples, Standard Methods for the Examination of Water and Wastewater recommends waiting 2 minutes for full color development. 13. Insert tube into chamber, close lid and select SCAN SAMPLE. Record result as ppm total chlorine. 14. SubtracttheFreeChlorinereadingfromtheTotalChlorinereadingto determine ppm combined chlorine. 15. Press OFF button to turn the colorimeter off or press EXIT button to exit to a previous menu or make another menu selection. NOTE: For best possible results, a reagent blank should be determined to account for any contribution to the test result by the reagent system. To determine the reagent blank, follow the above test procedure to scan a distilled or deionized water blank. Then follow the above procedure to perform the test on a distilled or deionized water sample. This test result is the reagent blank. Subtract the reagent blank from all subsequent test results of unknown samples. It is necessary to determine the reagent blank only when a new lot number of reagents is obtained. Smart2 TEST PROCEDURES 2.04 Chlorine- Liquid DPD 3/3
- Page 1: v.2.3 • 1.07 1919-test SMART 2Col
- Page 4 and 5: Range #of Test Factor (Test #) (ppm
- Page 7 and 8: ALKALINITY-UDV UNIT DOSE VIALS •
- Page 9 and 10: ALUMINUM ERIOCHROME CYANINE R METHO
- Page 11 and 12: AMMONIA-NITROGEN - LOW RANGE SALICY
- Page 13 and 14: PROCEDURE - SALT WATER 1. Press and
- Page 15 and 16: AMMONIA-NITROGEN - HIGH RANGE NESSL
- Page 17 and 18: BENZOTRIAZOLE/TOLYLTRIAZOLE UV Phot
- Page 19 and 20: BENZOTRIAZOLE PROCEDURE 1. Press an
- Page 21 and 22: BIGUANIDE COLORIMETRIC METHOD CODE
- Page 23 and 24: BORON AZOMETHINE-H METHOD CODE 486
- Page 25 and 26: BROMINE - UDV DPD METHOD-UNIT DOSE
- Page 27 and 28: PROCEDURE Use 10 mm square cell ada
- Page 29 and 30: CADMIUM PAN METHOD CODE 4017 QUANTI
- Page 31 and 32: CALCIUM & MAGNESIUM (TOTAL) HARDNES
- Page 33 and 34: CHLORIDE ARGENTOMETRIC METHOD CODE
- Page 35: CHLORINE LIQUID DPD METHOD CODE 48
- Page 39 and 40: CHLORINE-BROMINE-IODINE DPD METHOD
- Page 41 and 42: PROCEDURE-FREE CHLORINE 1. Press an
- Page 43 and 44: PROCEDURE A: BROMINE (NO CHLORINE)
- Page 45 and 46: IODINE Like chlorine and bromine, i
- Page 47 and 48: CHLORINE, FREE - UDV DPD METHOD-UNI
- Page 49 and 50: PROCEDURE Use 10 mm square cell ada
- Page 51 and 52: CHLORINE, TOTAL - UDV DPD METHOD-UN
- Page 53 and 54: PROCEDURE Use 10 mm square cell ada
- Page 55 and 56: CHLORINE DIOXIDE DPD METHOD CODE 36
- Page 57 and 58: CHROMIUM DIPHENYLCARBOHYDRAZIDE MET
- Page 59 and 60: CHROMIUM-HEXAVALENT DIPHENYLCARBOHY
- Page 61 and 62: COBALT PAN METHOD CODE 4851 QUANTI
- Page 63 and 64: COD-LOW RANGE MERCURY FREE DIGESTIO
- Page 65 and 66: PROCEDURE Use COD adapter (see p. 2
- Page 67 and 68: CHROMIUM-HEXAVALENT, TRIVALENT & TO
- Page 69 and 70: TOTAL CHROMIUM WITH ACID DIGESTION
- Page 71 and 72: COD-STANDARD RANGE MERCURY FREE DIG
- Page 73 and 74: PROCEDURE Use COD adapter (see p. 2
- Page 75 and 76: COD-HIGH RANGE MERCURY FREE DIGESTI
- Page 77 and 78: PROCEDURE Use COD adapter (see p. 2
- Page 79 and 80: COLOR PLATINUM COBALT METHOD NO RE
- Page 81 and 82: COPPER-LOW RANGE BICINCHONINIC ACID
- Page 83 and 84: COPPER CUPRIZONE METHOD CODE 4023 Q
- Page 85 and 86: COPPER DIETHYLDITHIOCARBAMATE METHO
SAMPLE<br />
HANDLING &<br />
PRESERVATION:<br />
Chlorine in aqueous solutions is not stable, and the chlorine<br />
content of samples or solutions, particularly weak solutions,<br />
will rapidly decrease. Exposure to sunlight or agitation will<br />
accelerate the reduction of chlorine present in such<br />
solutions. For best results, start analysis immediately after<br />
sampling. Samples to be analyzed for chlorine cannot be<br />
preserved or stored.<br />
INTERFERENCE: The only interfering substance likely to be encountered in<br />
water is oxidized manganese. The extent of this interference<br />
can be determined by treating a sample with sodium arsenite<br />
to destroy the chlorine present so that the degree of<br />
interference can be measured.<br />
Iodine and bromine can give a positive interference, but<br />
these are not normally present unless they have been added<br />
as sanitizers.<br />
Chlorine- Liquid DPD 2/3 Smart2 <strong>TEST</strong> PROCEDURES 2.04