PROCEDURE<br />

Use 10 mm square cell adapter.<br />

1. Press and hold ON button until colorimeter turns on.<br />

2. Press ENTER to start.<br />

3. Press ENTER to select <strong>TEST</strong>ING MENU.<br />

4. Select ALL <strong>TEST</strong>S (or another sequence containing 13 Ca&Mg H-UDV)<br />

from <strong>TEST</strong>ING MENU.<br />

5. Scroll to and selec t 13 Ca&Mg Hard-UDV from menu.<br />

6. Rinse a clean vial (0156) with sample water.<br />

7. Use the syringe (1184) to add 3 mL of sample to the vial.<br />

8. Insert the vial into chamber, close lid and select SCAN BLANK.<br />

9. Remove vial from the colorimeter.<br />

10. Use the syringe (1184) to add 3 mL of sample to a Calcium Hardness<br />

UDV vial (4309).<br />

11. Shake vigorously for 10 seconds.<br />

12. Wait one minute.<br />

13. Invert vial 3 times to mix.<br />

NOTE: Firmly tap side of vial 5-10 times to remove all air bubbles.<br />

14. Insert tube into chamber, close lid and select SCAN SAMPLE. Record<br />

result.<br />

15. Press OFF button to turn the colorimeter off or press EXIT button to exit<br />

to a previous menu or make another menu selection.<br />

NOTES: For best possible results, a reagent blank should be determined to<br />

account for any contribution to the test result by the reagent system. To<br />

determine the reagent blank, follow the above test procedure to scan a distilled<br />

or deionized water blank. Then follow the above procedure to perform the test<br />

on a distilled or deionized water sample. This test result is the reagent blank.<br />

Subtract the reagent blank from all subsequent test results of unknown samples.<br />

It is necessary to determine the reagent blank only when a new lot number of<br />

reagents are obtained.<br />

UDVs from opened pouches should be used promptly. Store unused vials from<br />

opened pouches in the Foil Storage Bag (9467) to extend the shelf life of the<br />

reagent. Generally, UDVs stored in the bag should be used within 10 days if<br />

the humidity is less than 50% and within 5 days if humidity is greater than<br />

50%. The Foil Storage Bag contains a desiccant pack with indicator. When<br />

the indicator in the window turns from blue to pink, the bag should be<br />

replaced.<br />

Calcium & Magnesium (Total) Hardness–UDV 2/2 Smart2 <strong>TEST</strong> PROCEDURES 2.04