BENZOTRIAZOLE PROCEDURE<br />

1. Press and hold ON button until colorimeter turns on.<br />

2. Press ENTER to start.<br />

3. Press ENTER to select Testing Menu.<br />

4. Select ALL <strong>TEST</strong>S (or another sequence containing 10 B triazole<br />

from <strong>TEST</strong>ING MENU.<br />

5. Scroll to and select 10 B triazole from menu.<br />

6. Rinse a tube (0290) with sample water. Fill to 10 mL with sample.<br />

7. Insert the tube into chamber, close lid and select SCAN BLANK.<br />

8. Remove the tube from colorimeter. Discard the sample.<br />

9. Adjust the sample water temperature to between 20 and 25C if necessary.<br />

10. Fill the graduated cylinder (0417) to the 25 mL line with sample water.<br />

Transfer to the Erlenmeyer Flask (2-2109).<br />

11. Use the pH Test Paper (9259) to check the pH of the sample. If the pH is<br />

not between 4 and 6, add one drop of *Sulfuric Acid, 1.0N (6139). Swirl<br />

to mix. Continue adding *Sulfuric Acid, 1.0N (6139) one drop at a time,<br />

swirling to mix and checking the pH after each drop, until the pH is<br />

between 4 and 6.<br />

12. Add 10 drops of NaK Tartrate (7841).<br />

13. Use the 0.25 g spoon (0695) to add one measure of *Benzotriazole<br />

Reagent (3818). Swirl to mix until the powder has dissolved.<br />

14. Replace the flask in the slot in the case. Insert the Penray Lamp (31041-1)<br />

into the flask. Plug in the Penray Power Source (31041-2) and turn the<br />

lamp on for exactly 5 minutes. Remove the lamp from the flask. Rinse and<br />

wipe the lamp dry.<br />

15. Fill a test tube (0290) to the 10 mL line with the digested sample. Cap<br />

tube.<br />

16. Insert tube into chamber, close lid and select SCAN SAMPLE. Record<br />

result in ppm Benzotriazole.<br />

17. Press OFF button to turn the colorimeter off or press EXIT button to exit<br />

to a previous menu or make another menu selection.<br />

NOTE: For best possible results, a reagent blank should be determined to<br />

account for any contribution to the test result by the reagent system. To<br />

determine the reagent blank, follow the above test procedure to scan a distilled<br />

or deionized water blank. Then follow the above procedure to perform the test<br />

on a distilled or deionized water sample. This test result is the reagent blank.<br />

Subtract the reagent blank from all subsequent test results of unknown samples.<br />

It is necessary to determine the reagent blank only when a new lot number of<br />

reagents is obtained.<br />