PROCEDURE<br />
Use COD adapter.<br />
1. Preheat COD reactor to 150 ±2°C. Follow safety precautions.<br />
2. Remove cap from a *Total Phosphorus Acid Reagent Tube (4035). Use a<br />
1.0 mL pipet (0354) to add 5.0 mL of sample.<br />
3. Use the 0.15 g spoon (0727) and a funnel (0459) to add one level measure<br />
of *Digestion Reagent Powder (4036). Tap funnel to dispense powder<br />
completely. Cap tube tightly and shake until powder completely dissolves.<br />
4. Place the tube in the COD reactor for 30 minutes.<br />
5. At the end of the heating period, turn the reactor off. Carefully remove<br />
the tube from the reactor and allow it to cool to room temperature.<br />
6. At the end of the cooling period, press and hold ON button until<br />
colorimeter turns on.<br />
7. Press ENTER to start.<br />
8. Press ENTER to select <strong>TEST</strong>ING MENU.<br />
9. Select ALL <strong>TEST</strong>S (or another sequence containing 82 Phos T LR)<br />
from <strong>TEST</strong>ING MENU.<br />
10. Scroll to and select 82 Phos T LR from the menu.<br />
11. Carefully remove the cap from the digested tube. Use another 1 mL pipet<br />
(0354) to add 1.0 mL of *Total Phosphorus LR Hydroxide Reagent (4038)<br />
to the tube. Cap and invert to mix.<br />
12. Wipe the tube with a damp towel to remove fingerprints and smudges.<br />
Wipe with a dry towel.<br />
13. Insert the tube into the chamber. Select SCAN BLANK. Remove the tube<br />
from the colorimeter.<br />
14. Use another 1 mL pipet (0354) to add *1.0 mL of Phosphate Acid<br />
Reagent (V-6282). Cap and invert tube to mix.<br />
15. Use the 0.1g spoon (0699) and a funnel (0459) to add one level spoon of<br />
Phosphate Reducing Reagent (V-6283). Tap funnel to dispense powder<br />
completely. Cap tube and shake until powder dissolves.<br />
16. Wait 5 minutes.<br />
17. Wipe the vials with a damp towel to remove fingerprints and smudges.<br />
Wipe with a dry towel.<br />
18. Insert the tube into the chamber. Select SCAN SAMPLE. Record the result<br />
as Total Phosphorus in mg/L PO4.<br />
19. Press OFF button to turn the colorimeter off or press EXIT button to exit<br />
to a previous menu or make another menu selection.<br />
NOTES: For greater accuracy, use laboratory grade pipets.<br />
To order reagent refills, Order Code R-4024.<br />
Smart2 <strong>TEST</strong> PROCEDURES 11.04 Phosphorus, Total–LR 3/3