INTERFERENCES: Sample color and turbidity interfere with the colorimetric pH measurement. Color interference may be removed bystandardizing the instrument with the original water sample. Two drops of 0.1N sodium thiosulfate per 100 mL of sample will eliminate chlorine interference. INDICATOR, RANGE, & TEST NAME: pH Indicator pH Smart2 Test Name Chlorphenol Red 5.0-6.8 74 pH CPR Phenol Red 6.6-8.4 75 pH PR Thymol Blue 8.0-9.6 76 pH TB pH 2/3 Smart2 TEST PROCEDURES 2.04
PROCEDURE 1. Use Indicator, Range, & Test Name chart to select the indicator, corresponding to anticipated pH range and to determine corresponding test name to select from colorimeter menu. 2. Press and hold ON button until colorimeter turns on. 3. Press ENTER to start. 4. Press ENTER to select TESTING MENU. 5. Select ALL TESTS (or another sequence containing the appropriate pH test name) from TESTING MENU. 6. Scroll to and select the appropriate pH test name from menu. 7. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample. 8. Insert tube into chamber, close lid and select SCAN BLANK. 9. Remove tube from colorimeter. Use the 0.5 mL pipet (0369) to add exactly 0.5 mL of the pH indicator for the chosen range. Cap and mix. 10. Insert tube into chamber, close lid and select SCAN SAMPLE. Record result. 11. Press OFF button to turn colorimeter off or press EXIT buttontoexittoa previous menu or make another menu selection. Smart2 TEST PROCEDURES 2.04 pH 3/3
- Page 107 and 108: HYDROGEN PEROXIDE-LOW RANGE DPD MET
- Page 109 and 110: HYDROGEN PEROXIDE- HIGH RANGE DPD M
- Page 111 and 112: HYDROGEN PEROXIDE-SHOCK DPD Method
- Page 113 and 114: IRON I,I0-PHENANTHROLINE METHOD COD
- Page 115 and 116: IRON BIPYRIDYL METHOD CODE 3648-SC
- Page 117 and 118: IRON-UDV BIPYRIDYL METHOD-UNIT DOSE
- Page 119 and 120: PROCEDURE Use 10 mm square cell ada
- Page 121 and 122: LEAD PAR METHOD CODE 4031 QUANTITY
- Page 123 and 124: MANGANESE-LOW RANGE PAN METHOD CODE
- Page 125 and 126: MANGANESE-HIGH RANGE PERIODATE METH
- Page 127 and 128: MERCURY TMK METHOD CODE 4861 QUANT
- Page 129 and 130: MOLYBDENUM-HIGH RANGE THIOGLYCOLATE
- Page 131 and 132: NICKEL DIMETHYLGLYOXIME METHOD CODE
- Page 133 and 134: NITRATE ZINC REDUCTION CODE 3689-S
- Page 135 and 136: NITRATE-NITROGEN-LOW RANGE CADMIUM
- Page 137 and 138: PROCEDURE NOTE: Place Dispenser Ca
- Page 139 and 140: NITRITE DIAZOTIZATION CODE 3694-SC
- Page 141 and 142: NITRITE-NITROGEN-LOW RANGE DIAZOTIZ
- Page 143 and 144: NITROGEN, TOTAL CHROMOTROPIC ACID W
- Page 145 and 146: PROCEDURE Use COD adapter. 1. Prehe
- Page 147 and 148: OXYGEN SCAVENGERS DEHA (Diethylhydr
- Page 149 and 150: CARBOHYDRAZIDE PROCEDURE 1. Press a
- Page 151 and 152: HYDROQUINONE PROCEDURE 1. Press and
- Page 153 and 154: OZONE INDIGO METHOD CODE 365I-SC QU
- Page 155 and 156: PROCEDURE-HIGH RANGE A. PREPARATION
- Page 157: pH COLORIMETRIC METHOD CODE 3700-SC
- Page 161 and 162: PHENOL AMINOANTIPYRINE METHOD CODE
- Page 163 and 164: PHOSPHATE-LOW RANGE ASCORBIC ACID R
- Page 165 and 166: PROCEDURE 1. Press and hold ON butt
- Page 167 and 168: PHOSPHATE-HIGH RANGE VANADOMOLYBDOP
- Page 169 and 170: PHOSPHORUS, TOTAL-LOW RANGE ASCORBI
- Page 171 and 172: PROCEDURE Use COD adapter. 1. Prehe
- Page 173 and 174: PHOSPHORUS, TOTAL-HIGH RANGE MOLYBD
- Page 175 and 176: PROCEDURE Use COD adapter. 1. Prehe
- Page 177 and 178: POTASSIUM TETRAPHENYLBORON METHOD C
- Page 179 and 180: SILICA-LOW RANGE HETEROPOLY BLUE ME
- Page 181 and 182: PROCEDURE 1. Press and hold ON butt
- Page 183 and 184: SILICA-HIGH RANGE SILICOMOLYBDATE M
- Page 185 and 186: SULFATE-HIGH RANGE BARIUM CHLORIDE
- Page 187 and 188: SULFIDE-LOW RANGE METHYLENE BLUE ME
- Page 189 and 190: SURFACTANTS ION PAIR EXTRACTION-BRO
- Page 191 and 192: TANNIN TUNGSTO-MOLYBDOPHOSPHORIC AC
- Page 193 and 194: TURBIDITY ABSORPTION METHOD NO REAG
- Page 195 and 196: ZINC-LOW RANGE ZINCON METHOD CODE 3
- Page 197 and 198: PROCEDURE A. PREPARATION OF DILUTE
- Page 199 and 200: Smart2 TEST PROCEDURES 2.04 SMART 2
- Page 201 and 202: APPENDIX Ammonia in water occurs in
PROCEDURE<br />
1. Use Indicator, Range, & Test Name chart to select the indicator,<br />
corresponding to anticipated pH range and to determine corresponding<br />
test name to select from colorimeter menu.<br />
2. Press and hold ON button until colorimeter turns on.<br />
3. Press ENTER to start.<br />
4. Press ENTER to select <strong>TEST</strong>ING MENU.<br />
5. Select ALL <strong>TEST</strong>S (or another sequence containing the appropriate pH<br />
test name) from <strong>TEST</strong>ING MENU.<br />
6. Scroll to and select the appropriate pH test name from menu.<br />
7. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with<br />
sample.<br />
8. Insert tube into chamber, close lid and select SCAN BLANK.<br />
9. Remove tube from colorimeter. Use the 0.5 mL pipet (0369) to add<br />
exactly 0.5 mL of the pH indicator for the chosen range. Cap and mix.<br />
10. Insert tube into chamber, close lid and select SCAN SAMPLE. Record<br />
result.<br />
11. Press OFF button to turn colorimeter off or press EXIT buttontoexittoa<br />
previous menu or make another menu selection.<br />
Smart2 <strong>TEST</strong> PROCEDURES 2.04 pH 3/3