24.04.2013 Views

TEST INSTRUCTIONS - Geotech Environmental Equipment

TEST INSTRUCTIONS - Geotech Environmental Equipment

TEST INSTRUCTIONS - Geotech Environmental Equipment

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

PROCEDURE<br />

1. Use the 10 mL graduated cylinder (0416) to measure 10 mL of sample<br />

water. Pour into glass test tube (0608).<br />

2. Use the 1 mL pipet (0354) to add 1 mL of *Hydrochloric Acid, 2.5N<br />

(6251).<br />

3. Use the 0.1 g spoon (0699) to add 2 measures of *Ammonium Persulfate<br />

Reagent (6566). Add two drops of *Silver Nitrate Solution, 0.0141N<br />

(6346WT). Mix until the powder has dissolved. The solution will be<br />

slightly cloudy at this point.<br />

4. Use 10 mL graduated cylinder (0416) to add 5 mL of Sodium Citrate, 10%<br />

(6253).<br />

5. Use a second 1 mL pipet (0354) to add 1 mL of *Ammonium Hydroxide,<br />

Conc. (6537). Mix, then dilute to 25 mL with deionized water.<br />

6. Use a third 1 mL pipet (0354) to add 1 mL of *Dimethylglyoxime, 1%<br />

(6254). Mix. Wait 20 minutes for color development.<br />

7. At end of 20 minute waiting period fill a clean tube (0290) to the 10 mL<br />

line with the developed test sample.<br />

8. Fill a second clean tube (0290) to 10 mL line with deionized water or<br />

untreated sample water. This is the blank.<br />

9. Press and hold ON button until colorimeter turns on.<br />

10. Press ENTER to start.<br />

11. Press ENTER to select <strong>TEST</strong>ING MENU.<br />

12. Select ALL <strong>TEST</strong>S (or another sequence containing 63 Nickel)from<br />

<strong>TEST</strong>ING MENU.<br />

13. Scroll to and select 63 Nickel from menu.<br />

14. Insert the blank into chamber, close lid and select SCAN BLANK.<br />

15. Insert test sample into chamber, close lid and select SCAN SAMPLE.<br />

Record result.<br />

16. Press OFF button to turn colorimeter off or press EXIT button to exit to a<br />

previous menu or make another menu selection.<br />

Nickel 2/2 Smart2 <strong>TEST</strong> PROCEDURES 2.04