PROCEDURE 1. Press and hold ON button until colorimeter turns on. 2. Press ENTER to start. 3. Press ENTER to select TESTING MENU. 4. Select ALL TESTS (or another sequence containing 54 Lead)from TESTING MENU. 5. Scroll to and select 54 Lead from menu. 6. Rinse a tube (0290) with sample water. Fill to 10 mL with sample. 7. Insert the tube into chamber, close lid and select SCAN BLANK. 8. Remove the tube from colorimeter. Use the Syringe (0807) to remove 5mL of sample from tube. Discard remaining sample. 9. Add the 5 mL of sample in the syringe to the tube. Add 5 mL Ammonium Chloride Buffer (4032) to fill the tube to the 10 mL line. Swirl to mix. 10. Add 3 drops *Sodium Cyanide, 10% (6565). Swirl to mix. 11. Use the 0.5 mL pipet (0353) to add 0.5 mL PAR Indicator (4033). Swirl to mix. 12. Use the 0.5 mL pipet (0353) to add 0.5 mL Stabilizing Reagent (4022). Cap and mix. 13. Insert tube into chamber, close lid and select SCAN SAMPLE. Record result in ppm as Reading A. 14. Remove tube from colorimeter. Add 3 drops DDC Reagent (4034). Cap and mix. 15. Insert tube into chamber, close lid and select SCAN SAMPLE. Record result in ppm as Reading B. 16. Calculate result: Lead (ppm) = Reading A Reading B 17. Press OFF button to turn the colorimeter off or press EXIT button to exit to a previous menu or make another menu selection. NOTE: For best possible results, a reagent blank should be determined to account for any contribution to the test result by the reagent system. To determine the reagent blank, follow the above test procedure to scan a distilled or deionized water blank. Then follow the above procedure to perform the test on a distilled or deionized water sample. This test result is the reagent blank. Subtract the reagent blank from all subsequent test results of unknown samples. It is necessary to determine the reagent blank only when a new lot number of reagents are obtained. Lead 2/2 Smart2 TEST PROCEDURES 2.04
MANGANESE–LOW RANGE PAN METHOD CODE 3658-01-SC QUANTITY CONTENTS CODE 4x30 mL *Hardness Buffer Reagent *4255-G 30 mL *Manganese Indicator Reagent *3956-G 15 mL *Sodium Cyanide, 10% *6565-E 1 Pipet, 0.5 mL, plastic 0369 1 Pipet, 1.0 mL, plastic 0354 *WARNING: Reagents marked with an * are considered hazardous substances. To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or our web site. To obtain a printed copy, contact us by e-mail, phone or fax. Manganese is present in ground water in the divalent state due to the lack of oxygen. In surface waters manganese may be in various oxidation states as soluble complexes or as suspended compounds. Manganese is rarely present in excess of 1 mg/L. It may cause an objectionable taste or cause staining problems in laundry, but manganese levels normally encountered in water seldom produce any health hazard. Manganese is removed from water by various means including chemical precipitation, pH adjustment, aeration, superchlorination and the use of ion exchange resins. APPLICATION: Drinking and surface waters; domestic and industrial wastewaters. RANGE: 0.00–0.70 ppm Manganese METHOD: PAN (1-2-Pyridylazo-2-Naphthol) forms a red complex SAMPLE HANDLING & PRESERVATION: with Manganese (Mn 2+) at a pH of 10 to 11. Manganese may oxidize readily in neutral water and precipitate from solution. It may adhere to or be absorbed by container walls, especially glass. Acidified samples can be stored in plastic. INTERFERENCES: None. Test is quite specific. Smart2 TEST PROCEDURES 2.04 Manganese–LR 1/2
- Page 71 and 72: COD-STANDARD RANGE MERCURY FREE DIG
- Page 73 and 74: PROCEDURE Use COD adapter (see p. 2
- Page 75 and 76: COD-HIGH RANGE MERCURY FREE DIGESTI
- Page 77 and 78: PROCEDURE Use COD adapter (see p. 2
- Page 79 and 80: COLOR PLATINUM COBALT METHOD NO RE
- Page 81 and 82: COPPER-LOW RANGE BICINCHONINIC ACID
- Page 83 and 84: COPPER CUPRIZONE METHOD CODE 4023 Q
- Page 85 and 86: COPPER DIETHYLDITHIOCARBAMATE METHO
- Page 87 and 88: COPPER-UDV BICINCHONINIC ACID METHO
- Page 89 and 90: PROCEDURE Use 10 mm square cell ada
- Page 91 and 92: CYANIDE PYRIDINE-BARBITURIC ACID ME
- Page 93 and 94: CYANURIC ACID MELAMINE METHOD-TURBI
- Page 95 and 96: CYANURIC ACID-UDV MELAMINE METHOD-T
- Page 97 and 98: DISSOLVED OXYGEN WINKLER COLORIMETR
- Page 99 and 100: PROCEDURE 1. Press and hold ON butt
- Page 101 and 102: FLUORIDE SPADNS METHOD CODE 3647-01
- Page 103 and 104: PROCEDURE 1. Press and hold ON butt
- Page 105 and 106: HYDRAZINE p-DIMETHYLAMINOBENZALDEHY
- Page 107 and 108: HYDROGEN PEROXIDE-LOW RANGE DPD MET
- Page 109 and 110: HYDROGEN PEROXIDE- HIGH RANGE DPD M
- Page 111 and 112: HYDROGEN PEROXIDE-SHOCK DPD Method
- Page 113 and 114: IRON I,I0-PHENANTHROLINE METHOD COD
- Page 115 and 116: IRON BIPYRIDYL METHOD CODE 3648-SC
- Page 117 and 118: IRON-UDV BIPYRIDYL METHOD-UNIT DOSE
- Page 119 and 120: PROCEDURE Use 10 mm square cell ada
- Page 121: LEAD PAR METHOD CODE 4031 QUANTITY
- Page 125 and 126: MANGANESE-HIGH RANGE PERIODATE METH
- Page 127 and 128: MERCURY TMK METHOD CODE 4861 QUANT
- Page 129 and 130: MOLYBDENUM-HIGH RANGE THIOGLYCOLATE
- Page 131 and 132: NICKEL DIMETHYLGLYOXIME METHOD CODE
- Page 133 and 134: NITRATE ZINC REDUCTION CODE 3689-S
- Page 135 and 136: NITRATE-NITROGEN-LOW RANGE CADMIUM
- Page 137 and 138: PROCEDURE NOTE: Place Dispenser Ca
- Page 139 and 140: NITRITE DIAZOTIZATION CODE 3694-SC
- Page 141 and 142: NITRITE-NITROGEN-LOW RANGE DIAZOTIZ
- Page 143 and 144: NITROGEN, TOTAL CHROMOTROPIC ACID W
- Page 145 and 146: PROCEDURE Use COD adapter. 1. Prehe
- Page 147 and 148: OXYGEN SCAVENGERS DEHA (Diethylhydr
- Page 149 and 150: CARBOHYDRAZIDE PROCEDURE 1. Press a
- Page 151 and 152: HYDROQUINONE PROCEDURE 1. Press and
- Page 153 and 154: OZONE INDIGO METHOD CODE 365I-SC QU
- Page 155 and 156: PROCEDURE-HIGH RANGE A. PREPARATION
- Page 157 and 158: pH COLORIMETRIC METHOD CODE 3700-SC
- Page 159 and 160: PROCEDURE 1. Use Indicator, Range,
- Page 161 and 162: PHENOL AMINOANTIPYRINE METHOD CODE
- Page 163 and 164: PHOSPHATE-LOW RANGE ASCORBIC ACID R
- Page 165 and 166: PROCEDURE 1. Press and hold ON butt
- Page 167 and 168: PHOSPHATE-HIGH RANGE VANADOMOLYBDOP
- Page 169 and 170: PHOSPHORUS, TOTAL-LOW RANGE ASCORBI
- Page 171 and 172: PROCEDURE Use COD adapter. 1. Prehe
PROCEDURE<br />
1. Press and hold ON button until colorimeter turns on.<br />
2. Press ENTER to start.<br />
3. Press ENTER to select <strong>TEST</strong>ING MENU.<br />
4. Select ALL <strong>TEST</strong>S (or another sequence containing 54 Lead)from<br />
<strong>TEST</strong>ING MENU.<br />
5. Scroll to and select 54 Lead from menu.<br />
6. Rinse a tube (0290) with sample water. Fill to 10 mL with sample.<br />
7. Insert the tube into chamber, close lid and select SCAN BLANK.<br />
8. Remove the tube from colorimeter. Use the Syringe (0807) to remove<br />
5mL of sample from tube. Discard remaining sample.<br />
9. Add the 5 mL of sample in the syringe to the tube. Add 5 mL Ammonium<br />
Chloride Buffer (4032) to fill the tube to the 10 mL line. Swirl to mix.<br />
10. Add 3 drops *Sodium Cyanide, 10% (6565). Swirl to mix.<br />
11. Use the 0.5 mL pipet (0353) to add 0.5 mL PAR Indicator (4033). Swirl<br />
to mix.<br />
12. Use the 0.5 mL pipet (0353) to add 0.5 mL Stabilizing Reagent (4022).<br />
Cap and mix.<br />
13. Insert tube into chamber, close lid and select SCAN SAMPLE. Record<br />
result in ppm as Reading A.<br />
14. Remove tube from colorimeter. Add 3 drops DDC Reagent (4034). Cap<br />
and mix.<br />
15. Insert tube into chamber, close lid and select SCAN SAMPLE. Record<br />
result in ppm as Reading B.<br />
16. Calculate result:<br />
Lead (ppm) = Reading A Reading B<br />
17. Press OFF button to turn the colorimeter off or press EXIT button to exit<br />
to a previous menu or make another menu selection.<br />
NOTE: For best possible results, a reagent blank should be determined to<br />
account for any contribution to the test result by the reagent system. To<br />
determine the reagent blank, follow the above test procedure to scan a distilled<br />
or deionized water blank. Then follow the above procedure to perform the test<br />
on a distilled or deionized water sample. This test result is the reagent blank.<br />
Subtract the reagent blank from all subsequent test results of unknown samples.<br />
It is necessary to determine the reagent blank only when a new lot number of<br />
reagents are obtained.<br />
Lead 2/2 Smart2 <strong>TEST</strong> PROCEDURES 2.04