www.wilamed.com<br />

<strong>OXIrate</strong><br />

<strong>Pulse</strong><br />

<strong>Oximeter</strong><br />

1275<br />

EN


<strong>OXIrate</strong> <strong>Pulse</strong> <strong>Oximeter</strong><br />

Portable pulse oximeter with patented Nellcor technology<br />

○ simple, intuitive operation<br />

○ with number, trend and chart display<br />

○ background lighting for visibility during the<br />

day and night<br />

○ Nellcor SpO 2 module finger clip sensor<br />

for adults<br />

○ temperature sensor<br />

○ battery backup for 48 hours<br />

○ intelligent optical and acoustic alarm system<br />

○ adjustable alarm volumes<br />

The portable <strong>OXIrate</strong> pulse oximeter is used to monitor functional arterial oxygen saturation (SpO 2 ),<br />

the pulse rate (PR) und body temperature (Temp). It is based on the latest OxiMax ® technology by<br />

Nellcor . The <strong>OXIrate</strong> pulse oximeter designed with corresponding sensor can be used for a diverse<br />

range of patients ranging from a neo-natal child to an adult in clinical or home care.<br />

<strong>OXIrate</strong> pulse oximeter exhibits intuitive operation based on its compact and well-arranged<br />

design. Unlike other portable pulse oximeters, <strong>OXIrate</strong> enables reliable monitoring even in<br />

difficult measurement conditions such as signal interferences or low perfusion.<br />

While in use, the <strong>OXIrate</strong> pulse oximeter records its results, which can be later exported to a<br />

computer for analysis, archiving and printing. This will assist the expert handling the device in<br />

acquiring an overview of the saturation. (Data storage is not possible in re-usable sensors due<br />

to data privacy reasons.)<br />

Features and specifications<br />

○ storage capacity for up to 100 patient data<br />

○ 300 hours of data storage<br />

○ data export for analysis, archiving and<br />

printing<br />

○ display of the last 10 record minutes as<br />

trend graphic of SpO 2 und PR<br />

○ practical silicon safety cover<br />

○ sturdy case with shoulder strap and belt<br />

fastener<br />

Classification<br />

Electricity supply Battery operation<br />

EMV classification Class B according to CISPR 11<br />

Application parts Protection class BF<br />

Protection class through body IP 12<br />

Operation type Long-term or spot measurement<br />

Measurement and weight<br />

Measurement 160 x 70 x 38 mm<br />

Weight 185 g (without batteries)


Ambient requirements<br />

Operation Transport and storage<br />

Temperature 5°C to +40°C -20°C to +55°C<br />

Relative humidity 25% - 80% (no condensation) 25% - 93% (no condensation)<br />

Air pressure 860 hPa - 1060 hPa 700 hPa - 1060 hPa<br />

Measurement scope<br />

SpO 2<br />

1% to 100%<br />

PR 20 to 300 bpm<br />

Temp 0°C to 50°C<br />

Accuracy<br />

SpO in 2<br />

Neonates 70 to 100% ± 3 LSD<br />

Adults 70 to 100% ± 2 LSD<br />

Low perfusion 70 to 100% ± 2 LSD<br />

PR in<br />

Neonates/Adults 20 to 300 spm ± 3 LSD<br />

Low perfusion 20 to 300 spm ± 3 LSD<br />

Temp 25°C to 45°C ± 0.1°C<br />

0°C to 25°C and 45°C to 50°C ± 0.2°C<br />

Display<br />

○ LCD matrix with 128 x 64 pixels with<br />

LED background lighting<br />

○ Time of data update 2 s<br />

Batteries<br />

○ Standard design 4 x 1.5 V AA alkaline<br />

batteries<br />

○ Typical life of batteries 48 hours<br />

○ Also usable 4 rechargeable batteries<br />

1.2 V Ni-H<br />

○ Charging of built-in batteries is not possible<br />

Function<br />

○ Alarm messages are sent acoustically and<br />

optically<br />

○ Acoustic Alarms can be set in silence mode<br />

for duration of 30, 60, 90 and 120 seconds<br />

Data storage<br />

○ Log with up to 300 hours of recording<br />

for SpO 2 and PR<br />

○ Storage for 100 patient data (sequenced<br />

according to identification number)<br />

Activation<br />

SpO 2<br />

1%<br />

PR 1 bpm<br />

Temp 0,1 °C<br />

Trend display<br />

○ Trend graphic of the last 10 record minutes<br />

of SpO2 and PR<br />

Scope of supply<br />

○ Reusable Nellcor finger sensor 100.620<br />

for adults, model Nellcor<br />

DS-100A Durasensor ®<br />

(weak perfusion resistance)<br />

○ Temperature sensor 100.621<br />

○ User manual 100.622<br />

○ Carry case 100.623<br />

○ Batteries<br />

Accessories<br />

○ Technical instructions 100.624<br />

○ Silicon safety cover 100.625<br />

○ Battery charger 100.626<br />

○ Patient data software (includes 100.627<br />

1 CD, USB cable, User manual)


<strong>WILAmed</strong> GmbH<br />

Medizinische Geräte und Zubehör<br />

Aurachhöhe 5–7<br />

Gewerbepark Barthelmesaurach<br />

91126 Kammerstein<br />

Germany<br />

Telephone: +49(0)9178 / 99 69 99-0<br />

Facsimile: +49(0)9178 / 99 67 78<br />

E-Mail: info@wilamed.com<br />

www.wilamed.com<br />