<strong>OXIrate</strong><br />
<strong>Pulse</strong><br />
<strong>Oximeter</strong><br />
1275<br />
EN
<strong>OXIrate</strong> <strong>Pulse</strong> <strong>Oximeter</strong><br />
Portable pulse oximeter with patented Nellcor technology<br />
○ simple, intuitive operation<br />
○ with number, trend and chart display<br />
○ background lighting for visibility during the<br />
day and night<br />
○ Nellcor SpO 2 module finger clip sensor<br />
for adults<br />
○ temperature sensor<br />
○ battery backup for 48 hours<br />
○ intelligent optical and acoustic alarm system<br />
○ adjustable alarm volumes<br />
The portable <strong>OXIrate</strong> pulse oximeter is used to monitor functional arterial oxygen saturation (SpO 2 ),<br />
the pulse rate (PR) und body temperature (Temp). It is based on the latest OxiMax ® technology by<br />
Nellcor . The <strong>OXIrate</strong> pulse oximeter designed with corresponding sensor can be used for a diverse<br />
range of patients ranging from a neo-natal child to an adult in clinical or home care.<br />
<strong>OXIrate</strong> pulse oximeter exhibits intuitive operation based on its compact and well-arranged<br />
design. Unlike other portable pulse oximeters, <strong>OXIrate</strong> enables reliable monitoring even in<br />
difficult measurement conditions such as signal interferences or low perfusion.<br />
While in use, the <strong>OXIrate</strong> pulse oximeter records its results, which can be later exported to a<br />
computer for analysis, archiving and printing. This will assist the expert handling the device in<br />
acquiring an overview of the saturation. (Data storage is not possible in re-usable sensors due<br />
to data privacy reasons.)<br />
Features and specifications<br />
○ storage capacity for up to 100 patient data<br />
○ 300 hours of data storage<br />
○ data export for analysis, archiving and<br />
printing<br />
○ display of the last 10 record minutes as<br />
trend graphic of SpO 2 und PR<br />
○ practical silicon safety cover<br />
○ sturdy case with shoulder strap and belt<br />
fastener<br />
Classification<br />
Electricity supply Battery operation<br />
EMV classification Class B according to CISPR 11<br />
Application parts Protection class BF<br />
Protection class through body IP 12<br />
Operation type Long-term or spot measurement<br />
Measurement and weight<br />
Measurement 160 x 70 x 38 mm<br />
Weight 185 g (without batteries)
Ambient requirements<br />
Operation Transport and storage<br />
Temperature 5°C to +40°C -20°C to +55°C<br />
Relative humidity 25% - 80% (no condensation) 25% - 93% (no condensation)<br />
Air pressure 860 hPa - 1060 hPa 700 hPa - 1060 hPa<br />
Measurement scope<br />
SpO 2<br />
1% to 100%<br />
PR 20 to 300 bpm<br />
Temp 0°C to 50°C<br />
Accuracy<br />
SpO in 2<br />
Neonates 70 to 100% ± 3 LSD<br />
Adults 70 to 100% ± 2 LSD<br />
Low perfusion 70 to 100% ± 2 LSD<br />
PR in<br />
Neonates/Adults 20 to 300 spm ± 3 LSD<br />
Low perfusion 20 to 300 spm ± 3 LSD<br />
Temp 25°C to 45°C ± 0.1°C<br />
0°C to 25°C and 45°C to 50°C ± 0.2°C<br />
Display<br />
○ LCD matrix with 128 x 64 pixels with<br />
LED background lighting<br />
○ Time of data update 2 s<br />
Batteries<br />
○ Standard design 4 x 1.5 V AA alkaline<br />
batteries<br />
○ Typical life of batteries 48 hours<br />
○ Also usable 4 rechargeable batteries<br />
1.2 V Ni-H<br />
○ Charging of built-in batteries is not possible<br />
Function<br />
○ Alarm messages are sent acoustically and<br />
optically<br />
○ Acoustic Alarms can be set in silence mode<br />
for duration of 30, 60, 90 and 120 seconds<br />
Data storage<br />
○ Log with up to 300 hours of recording<br />
for SpO 2 and PR<br />
○ Storage for 100 patient data (sequenced<br />
according to identification number)<br />
Activation<br />
SpO 2<br />
1%<br />
PR 1 bpm<br />
Temp 0,1 °C<br />
Trend display<br />
○ Trend graphic of the last 10 record minutes<br />
of SpO2 and PR<br />
Scope of supply<br />
○ Reusable Nellcor finger sensor 100.620<br />
for adults, model Nellcor<br />
DS-100A Durasensor ®<br />
(weak perfusion resistance)<br />
○ Temperature sensor 100.621<br />
○ User manual 100.622<br />
○ Carry case 100.623<br />
○ Batteries<br />
Accessories<br />
○ Technical instructions 100.624<br />
○ Silicon safety cover 100.625<br />
○ Battery charger 100.626<br />
○ Patient data software (includes 100.627<br />
1 CD, USB cable, User manual)
<strong>WILAmed</strong> GmbH<br />
Medizinische Geräte und Zubehör<br />
Aurachhöhe 5–7<br />
Gewerbepark Barthelmesaurach<br />
91126 Kammerstein<br />
Germany<br />
Telephone: +49(0)9178 / 99 69 99-0<br />
Facsimile: +49(0)9178 / 99 67 78<br />
E-Mail: info@wilamed.com<br />
www.wilamed.com<br />
Version 1.0<br />
Status: 16.03.2009