Operating Instructions Metriso 2000 - Test Kit
Operating Instructions Metriso 2000 - Test Kit
Operating Instructions Metriso 2000 - Test Kit
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>Metriso</strong> <strong>2000</strong> - <strong>Operating</strong> <strong>Instructions</strong><br />
Part No. 7100.<strong>2000</strong>.MK<br />
Measuring<br />
Select the RESD menu and a test voltage of U = 10 V. If a value of less than 100 kΩ is displayed,<br />
the displayed value is the result. If R > 100 KΩ is displayed, the test voltage must be changed to<br />
100 volts. The displayed value then, is the result.<br />
Surface Resistance RS<br />
Probe<br />
Model 850<br />
Probe Model 880<br />
Insulative Plate<br />
Surface Resistance RPP<br />
Probe<br />
Model 850<br />
Volume Resistance RV<br />
Probe<br />
Model 850<br />
Probe Model 880<br />
Insulative Plate<br />
Metal counter-electrode<br />
blue<br />
red<br />
sblack<br />
red black<br />
blue<br />
black<br />
red<br />
Probe<br />
Model 850<br />
Standard:<br />
DIN EN 61340-2-3<br />
Standard:<br />
DIN EN 61340-2-3<br />
Standard:<br />
DIN EN 61340-2-3<br />
Page 3 of 9 V0708<br />