<strong>Metriso</strong> <strong>2000</strong> - <strong>Operating</strong> <strong>Instructions</strong><br />

Part No. 7100.<strong>2000</strong>.MK<br />

Measuring<br />

Select the RESD menu and a test voltage of U = 10 V. If a value of less than 100 kΩ is displayed,<br />

the displayed value is the result. If R > 100 KΩ is displayed, the test voltage must be changed to<br />

100 volts. The displayed value then, is the result.<br />

Surface Resistance RS<br />

Probe<br />

Model 850<br />

Probe Model 880<br />

Insulative Plate<br />

Surface Resistance RPP<br />

Probe<br />

Model 850<br />

Volume Resistance RV<br />

Probe<br />

Model 850<br />

Probe Model 880<br />

Insulative Plate<br />

Metal counter-electrode<br />

blue<br />

red<br />

sblack<br />

red black<br />

blue<br />

black<br />

red<br />

Probe<br />

Model 850<br />

Standard:<br />

DIN EN 61340-2-3<br />

Standard:<br />

DIN EN 61340-2-3<br />

Standard:<br />

DIN EN 61340-2-3<br />