Alarms<br />

Alarm Symbol Ca<strong>use</strong> Remedy<br />

Battery alarm in<br />

mains mode<br />

Plug symbol visible in display<br />

22 4137-2_nu_Applix_<strong>Smart</strong>_GB<br />

GB<br />

Batteries are defective. In<strong>for</strong>m personnel who<br />

service pumps.<br />

Batteries should be changed<br />

by authorised engineers<br />

only!<br />

Plug symbol not visible in display<br />

Mains control indicator on<br />

pump holder does not<br />

come on.<br />

Faulty mains connection.<br />

Mains control indicator on<br />

pump holder comes on.<br />

<strong>Pump</strong> is not properly<br />

inserted in pump holder.<br />

Contact pins of pump and<br />

pump holder are dirty.<br />

Check that mains cable is<br />

properly fixed in position and<br />

is not damaged in any way.<br />

Insert pump in pump holder.<br />

Ensure it has clicked in to<br />

place.<br />

Remove dirt with cloth and<br />

soapy water or as directed<br />

by hospital policy (see p. 28).<br />

Allow the pump to dry.