APPLIX Smart - Enteral Feeding Pump - Instructions for use
APPLIX Smart - Enteral Feeding Pump - Instructions for use
APPLIX Smart - Enteral Feeding Pump - Instructions for use
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Alarms<br />
Alarm Symbol Ca<strong>use</strong> Remedy<br />
Battery alarm in<br />
mains mode<br />
Plug symbol visible in display<br />
22 4137-2_nu_Applix_<strong>Smart</strong>_GB<br />
GB<br />
Batteries are defective. In<strong>for</strong>m personnel who<br />
service pumps.<br />
Batteries should be changed<br />
by authorised engineers<br />
only!<br />
Plug symbol not visible in display<br />
Mains control indicator on<br />
pump holder does not<br />
come on.<br />
Faulty mains connection.<br />
Mains control indicator on<br />
pump holder comes on.<br />
<strong>Pump</strong> is not properly<br />
inserted in pump holder.<br />
Contact pins of pump and<br />
pump holder are dirty.<br />
Check that mains cable is<br />
properly fixed in position and<br />
is not damaged in any way.<br />
Insert pump in pump holder.<br />
Ensure it has clicked in to<br />
place.<br />
Remove dirt with cloth and<br />
soapy water or as directed<br />
by hospital policy (see p. 28).<br />
Allow the pump to dry.