PROTOCOL CONTINUED FROM INSIDE FRONT COVER<br />

Washing and Draining Gently layer on 750µl ice-cold Acid-Salt Wash<br />

Reagent to the collagen-dye pellet to remove<br />

unbound dye from the surface of the pellet and the<br />

inside surface of the microcentrifuge tube.<br />

Centrifuge at 12,000 r.p.m. for 10 minutes.<br />

Drain the wash into a waste container and carefully<br />

remove any fluid from the lip of the tubes using<br />

cotton wool buds.<br />

Release and Recovery of Add 250µl Alkali Reagent to reagent blanks,<br />

Collagen Bound Dye standards and samples.<br />

Recap tubes and release the collagen bound dye<br />

into solution. A vortex mixer is suitable.<br />

When all of the bound dye has been dissolved,<br />

usually within 5 minutes, the samples are ready for<br />

measurement. The colour is light stable, but<br />

should be read within 2 to 3 hours. Keep tubes<br />

capped until ready to measure Absorbance.<br />

Measurement Transfer 200µl of each sample to individual wells of<br />

a 96 micro well plate, (keep a record map of the<br />

contents of each well; A1 to H12).<br />

Set the microplate reader to 555nm, or the closest<br />

matching blue-green colour filter.<br />

Measure absorbance against water for the reagent<br />

blanks, standards and test samples.<br />

Obtain collagen concentrations from the Standard<br />

Curve. Duplicates should be close to + 5% of their<br />

mean value.<br />

If sample absorbance values are at the top end of the standard curve the assay<br />

should be repeated and a set of higher concentration standards (15, 30 and 50µg)<br />

used, together with 1000µl of alkali reagent rather than 250µl. The assay is<br />

completed as above.<br />

READ SIRCOL MANUAL FOR SAMPLE PREPARATION DETAILS<br />