<strong>Operative</strong> <strong>Technique</strong> – Retrograde <strong>Technique</strong><br />
Repositioning may be carried out<br />
either by hand or by attaching the<br />
Universal Rod, Short to the Nail<br />
Holding Screw. The slotted hammer<br />
may be used to reposition the nail<br />
smoothly (Fig. 62). DO NOT hit on<br />
the Target Device.<br />
When locking the retrograde nail in<br />
the Static Mode, the nail is countersunk<br />
a minimum of 6mm below the<br />
surface. When the implant is inserted<br />
in the Dynamic Mode, with active<br />
apposition/compression, or in the<br />
Advanced Locking Mode, the recommended<br />
insertion depth is 10mm.<br />
Note :<br />
Remove the Guide Wire prior to<br />
drilling and inserting the Locking<br />
Screws.<br />
34<br />
Fig. 61<br />
Fig. 62