<strong>Operative</strong> <strong>Technique</strong> – Antegrade <strong>Technique</strong><br />
Apposition/Compression Locking Mode<br />
In transverse or axially stable fracture<br />
patterns, active apposition/compression<br />
increases fracture stability and<br />
enhances fracture healing. The antegrade<br />
<strong>T2</strong> <strong>Humeral</strong> Nail provides the<br />
option to treat a humerus fracture<br />
with active mechanical apposition/<br />
compression prior to leaving the<br />
operating room.<br />
Note:<br />
Distal freehand static locking<br />
must be performed prior to<br />
applying active, controlled apposition/compression<br />
to the fracture<br />
site.<br />
If active apposition/compression is<br />
required, a Partially Threaded Locking<br />
Screw (Shaft Screw) is inserted via the<br />
Target Device in the dynamic position<br />
of the oblong hole. This will allow for<br />
a maximum of 6mm of active, controlled<br />
apposition/compression. In<br />
order to insert the Partially Threaded<br />
Locking Screw (Shaft Screw), drill<br />
both cortices with the Ø3.5 × 230mm<br />
Drill (1806-3540S). Next, the near<br />
cortex ONLY is overdrilled with the<br />
Ø4.0 × 180mm Drill (1806-4000S).<br />
Note:<br />
After the opposite cortex is drilled<br />
with the Ø3.5 × 230mm drill, the<br />
correct screw length can be read<br />
directly off of the calibrated Drill<br />
at the end of the Drill Sleeve.<br />
After the Partially Threaded Locking<br />
Screw (Shaft Screw) is inserted, the<br />
Nail Holding Screw is removed,<br />
leaving the insertion post intact with<br />
the nail (Fig. 35). This will act as a<br />
guide for the Compression Screw. The<br />
Compression Screw with the Compression<br />
Screwdriver Shaft (1806-0263)<br />
assembled on the Teardrop Handle is<br />
inserted through the insertion post<br />
(Fig. 36).<br />
25<br />
Fig. 35<br />
Fig. 36