340<br />

<strong>Biomedical</strong> <strong>Engineering</strong> <strong>–</strong> <strong>From</strong> <strong>Theory</strong> <strong>to</strong> <strong>Applications</strong><br />

The same inven<strong>to</strong>r has also obtained a patent in respect of another, more recent but very<br />

different design. The relevant patent is CN2009/000406. It is unclear whether the later<br />

design is intended <strong>to</strong> supersede the original, or compete with it.<br />

10.6 SmartKlamp<br />

Inven<strong>to</strong>r: Ten Have, H.F.<br />

Primary patent: MY0100375<br />

Patent priority date: 30 Jan 2001<br />

Patient age range: Full-term neonate <strong>to</strong> mid-puberty<br />

Website: http://smartcircumcision.com/index.html<br />

Category: Ischaemic necrosis device<br />

Fig. 14. The SmartKlamp.<br />

Procedure: The SmartKlamp is an ischaemic necrosis device functioning in broadly the same<br />

way as the Ali's Klamp, Ismail Klamp and Tara KLamp.<br />

Since its invention by Henri Ferdinand Ten Have, the commercial his<strong>to</strong>ry of the<br />

SmartKlamp has <strong>to</strong> some extent interfered with its development and availability. Originally<br />

produced only in child sizes by the Dutch company Circumvent BV (Hengelo, The<br />

Netherlands), supply ceased when Circumvent BV went in<strong>to</strong> liquidation, having apparently<br />

over-spent on research and development of the adult model. It should perhaps be noted that<br />

Circumvent's experimental adult device had an angled clamping ring similar in principle <strong>to</strong><br />

that now found in the Ali's Klamp.<br />

The apparently identical product in child sizes has now appeared in Malaysia, marketed by<br />

a supplier called Smartcircumcision. Aldemir and co-workers described the device thus:<br />

"This fits on the penis much as the others do. After 4 days the connection between its inner<br />

tube and casing is cut and removed. The inner tube is then left <strong>to</strong> fall off spontaneously in<br />

time. Median operative time is 8 minutes, compared with 18 minutes for conventional<br />

dissection, and cosmetic result, judged blinded by a urologist, was better. Parents'<br />

satisfaction scores were the same" (Aldemir et al., 2008).<br />

The SmartKlamp differs from the Ali's Klamp only in one significant respect: The clamping<br />

ring is not angled <strong>to</strong> match the typical slant of the coronal rim of the glans. In theory this<br />

achieves less satisfac<strong>to</strong>ry capture of inner foreskin ventrally, but the original manufacturer<br />

claimed it is possible <strong>to</strong> adjust the foreskin <strong>to</strong> compensate. The SmartKlamp apparently<br />

shares with the Ali's Klamp one notable design weakness. An unwilling or meddlesome<br />

patient could, in theory, open the clamp by placing a flat object such as a table knife between<br />

the tube and the locking arms and then twisting. Additionally, such action would render the