Maintenance Log<br />

S E C T I O N 1 0 • M a i n t e n a n c e<br />

All maintenance and testing should be done by competent personnel.<br />

Regularly scheduled maintenance ensures proper functioning of the<br />

equipment. Refer to the table below for required tasks and frequency of<br />

routine maintenance.<br />

Maintenance Checklist<br />

Task Every Use Every 30 Days Every 12 Months<br />

Clean the Exterior <br />

General Inspection <br />

Disinfect the Reservoir and<br />

Change Distilled Water or<br />

Isopropyl Alcohol solution<br />

Lubricate the O-Rings <br />

Disinfect the Reservoir and<br />

Change the Hydrogen<br />

Peroxide Solution<br />

Replace the O-Rings <br />

Alarm Signal Test <br />

Add Recirculating Solution<br />

Test<br />

Check Disposables Test <br />

Over Temperature Alarm<br />

Test<br />

Verify Temperature<br />

Calibration<br />

Electrical Safety Tests <br />

HOTLINE ® Blood and Fluid Warmer | Operator’s <strong>Manual</strong> 33