Maintenance Log<br />
S E C T I O N 1 0 • M a i n t e n a n c e<br />
All maintenance and testing should be done by competent personnel.<br />
Regularly scheduled maintenance ensures proper functioning of the<br />
equipment. Refer to the table below for required tasks and frequency of<br />
routine maintenance.<br />
Maintenance Checklist<br />
Task Every Use Every 30 Days Every 12 Months<br />
Clean the Exterior <br />
General Inspection <br />
Disinfect the Reservoir and<br />
Change Distilled Water or<br />
Isopropyl Alcohol solution<br />
Lubricate the O-Rings <br />
Disinfect the Reservoir and<br />
Change the Hydrogen<br />
Peroxide Solution<br />
Replace the O-Rings <br />
Alarm Signal Test <br />
Add Recirculating Solution<br />
Test<br />
Check Disposables Test <br />
Over Temperature Alarm<br />
Test<br />
Verify Temperature<br />
Calibration<br />
Electrical Safety Tests <br />
HOTLINE ® Blood and Fluid Warmer | Operator’s <strong>Manual</strong> 33