Periodic Electrical Testing<br />

S E C T I O N 9 • T e s t i n g<br />

Electrical Safety Tests must be performed by competent personnel<br />

authorized by the institution to perform such testing. The Safety Tests<br />

must be performed and documented at least once per year, or according to<br />

institutional policy. These tests include but are not limited to:<br />

• Leakage current<br />

• Ground bond test<br />

Leakage Current<br />

Leakage current must be tested according to methods and pass/fail<br />

criteria described in UL 2601-1 or EN 60601-1. Leakage current must be<br />

performed with the heater circuit in the full ON condition. To achieve<br />

this condition, perform the test when the reservoir is at room<br />

temperature. When the HOTLINE ® Warmer is first turned on and the<br />

temperature is rapidly rising, but still below 41°C, the heater circuit is in<br />

a full ON condition.<br />

Note: The HOTLINE ® Warmer is equipped with sensing interlocks. A<br />

HOTLINE ® Fluid Warming Set is required to correctly operate the<br />

HOTLINE ® Warmer and perform leakage current testing. Do not defeat<br />

the sensing interlocks or try to operate the HOTLINE ® without a<br />

HOTLINE ® Fluid Warming Set in place.<br />

Ground Bond Test<br />

Ground bond test must be tested according to methods and pass/fail<br />

criteria described in UL 2601-1 or EN 60601-1.<br />

HOTLINE ® Blood and Fluid Warmer | Operator’s <strong>Manual</strong> 27