The HOTLINE ® Warmer should be tested by hospital biomedical<br />

personnel prior to placing it in service. All testing and maintenance should<br />

be performed by competent personnel. If competent personnel are not<br />

available, contact <strong>Smiths</strong> <strong>Medical</strong> or your local <strong>Smiths</strong> <strong>Medical</strong> distributor.<br />

If the HOTLINE ® Warmer and any installed accessories do not pass any of<br />

the listed tests, discontinue use of the HOTLINE ® Warmer and remove<br />

from service. Contact <strong>Smiths</strong> <strong>Medical</strong> or your local <strong>Smiths</strong> <strong>Medical</strong><br />

distributor.<br />

WARNINGS<br />

• If any visual indicator does not illuminate or the audible signal<br />

does not sound, do not use the Fluid Warmer. Remove the<br />

device from service immediately. Death or serious injury may<br />

occur to the patient or user if this warning is not followed.<br />

Note: Alarm testing requires a HOTLINE ® Fluid Warming Set to be<br />

installed and that the HOTLINE ® Warmer be turned ON and in the<br />

Operating mode.<br />

Alarm Signal Test<br />

The Alarm Signal Test is used to confirm proper operation of the visual<br />

and audible alarm indicators.<br />

1 Press and hold the Alarm Test button (a).<br />

2 Observe the following:<br />

• The green Operating LED turns off.<br />

• Three red LEDs (Check Disposables, Add Solution, and Over<br />

Temperature) illuminate.<br />

• The audible alarm sounds and repeats approximately every two<br />

seconds.<br />

HOTLINE ® Blood and Fluid Warmer | Operator’s <strong>Manual</strong> 23