S E C T I O N 4 • I m p o r t a n t S a f e t y I n f o r m a t i o n<br />

WARNINGS<br />

Continued from previous page<br />

• The Fluid Warmer is for use only with <strong>Smiths</strong> <strong>Medical</strong> supplied<br />

or approved parts, accessories, and Irrigation Disposable Sets.<br />

The device may not function as intended with the use of<br />

unapproved parts, accessories, or Disposable Sets.<br />

• Grounding reliability can only be achieved when MAINS power<br />

cords are connected to a properly grounded receptacle. Risk of<br />

electrical shock exists if the equipment is not connected to a<br />

properly grounded receptacle.<br />

• If any visual indicator does not illuminate or the audible signal<br />

does not sound, do not use the Fluid Warmer. Remove the<br />

device from service immediately.<br />

• Do not operate the Fluid Warmer in the presence of a<br />

flammable anesthetic mixture with air, oxygen, or nitrous<br />

oxide.The risk of explosion exists if the Fluid Warmer is operated<br />

in a potentially explosive environment.<br />

• No user-serviceable parts. All service must be performed by<br />

<strong>Smiths</strong> <strong>Medical</strong> or an authorized representative.<br />

• Disposable Sets are supplied with a sterile fluid path which may<br />

be compromised if the caps are not in place. Do not use<br />

Disposable Set if spike caps and end caps are not securely in<br />

place, or if connections are not secure as the fluid path may not<br />

be sterile and may cause death or serious injury to the patient.<br />

• Disposable Sets are for single use only. To reduce the risk of<br />

cross contamination, do not reuse Disposable Sets.<br />

• Do not overextend Automatic Pole; if Automatic Pole is<br />

overextended the Fluid Warmer may become unstable.<br />

Level 1 ® NORMOFLO ® H-<strong>1100</strong> Irrigation Fluid Warmer | Operator’s Manual 5