EQUATOR® Convective Warmer - Smiths Medical
EQUATOR® Convective Warmer - Smiths Medical
EQUATOR® Convective Warmer - Smiths Medical
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
SECTION 12<br />
Specifications<br />
System Specifications<br />
Standard Compliance Guidelines<br />
Product Safety EN 60601-1, UL 2601-1<br />
<strong>Convective</strong> Warming Blanket Flammability NFPA 702, 16 CFR 1610<br />
EMC EN 60601-1-2, FCC 47 CFR Part 15, Class B<br />
Enclosure Protection IEC 60529 IP Code: IPX1<br />
<strong>Convective</strong> <strong>Warmer</strong>s ASTM F2196-02<br />
Drop Test Compliance EN 60601-1<br />
Physical Dimensions<br />
Height, Overall 30 cm (11.75 inches)<br />
Width, Overall 24 cm (9.5 inches)<br />
Depth, Overall 19 cm (7.5 inches)<br />
Weight, Overall 6.8 Kg (15 lbs)<br />
Air Flow 8.4 to 12.7m/sec (1650 to 2500 feet/min)<br />
1.59 to 2.41 cm/min (56.2 to 85.2 cf/min)<br />
Filtration System 0.2 micron filter<br />
Maximum Height on I.V. Pole 117 cm (46 inches)<br />
S E C T I O N 1 2 • S p e c i f i c a t i o n s<br />
Shipping Durability Vibration and Drop Compliant with ISTA 1A and 2A<br />
Environmental Temperature Humidity<br />
Operation 10 o C to 40 o C 10 to 95% (non condensing)<br />
Transportation -18 o C to 60 o C 5 to 95% (non condensing)<br />
Storage 5 o C to 40 o C 5 to 95% (non condensing)<br />
Thermal Temperature<br />
Hose End Temperature 36° ± 1°C (97°F ± 2°F)<br />
40° ± 1°C (104°F ± 2°F)<br />
44° ± 1°C (111°F ± 2°F)<br />
Over Temperature Alarm - Forced Air Over<br />
Temperature Protection<br />
36°C = 39° ± 1°C (102°F ± 2°F)<br />
40°C = 43° ± 1°C (109°F ± 2°F)<br />
44°C = 47° ± 1°C (117°F ± 2°F)<br />
Electrical<br />
Supply Power Input:<br />
Type<br />
100V 100 VAC, 50/60 Hz, 9.00 Amps<br />
115V 115 VAC, 50/60 Hz, 8.05 Amps<br />
230V 230 VAC, 50/60 Hz, 4.0 Amps<br />
Protection Against Electrical Shock Class 1 Equipment<br />
Type BF<br />
EQUATOR ® <strong>Convective</strong> <strong>Warmer</strong> | Operator’s Manual 33