SECTION 12<br />

Specifications<br />

System Specifications<br />

Standard Compliance Guidelines<br />

Product Safety EN 60601-1, UL 2601-1<br />

<strong>Convective</strong> Warming Blanket Flammability NFPA 702, 16 CFR 1610<br />

EMC EN 60601-1-2, FCC 47 CFR Part 15, Class B<br />

Enclosure Protection IEC 60529 IP Code: IPX1<br />

<strong>Convective</strong> <strong>Warmer</strong>s ASTM F2196-02<br />

Drop Test Compliance EN 60601-1<br />

Physical Dimensions<br />

Height, Overall 30 cm (11.75 inches)<br />

Width, Overall 24 cm (9.5 inches)<br />

Depth, Overall 19 cm (7.5 inches)<br />

Weight, Overall 6.8 Kg (15 lbs)<br />

Air Flow 8.4 to 12.7m/sec (1650 to 2500 feet/min)<br />

1.59 to 2.41 cm/min (56.2 to 85.2 cf/min)<br />

Filtration System 0.2 micron filter<br />

Maximum Height on I.V. Pole 117 cm (46 inches)<br />

S E C T I O N 1 2 • S p e c i f i c a t i o n s<br />

Shipping Durability Vibration and Drop Compliant with ISTA 1A and 2A<br />

Environmental Temperature Humidity<br />

Operation 10 o C to 40 o C 10 to 95% (non condensing)<br />

Transportation -18 o C to 60 o C 5 to 95% (non condensing)<br />

Storage 5 o C to 40 o C 5 to 95% (non condensing)<br />

Thermal Temperature<br />

Hose End Temperature 36° ± 1°C (97°F ± 2°F)<br />

40° ± 1°C (104°F ± 2°F)<br />

44° ± 1°C (111°F ± 2°F)<br />

Over Temperature Alarm - Forced Air Over<br />

Temperature Protection<br />

36°C = 39° ± 1°C (102°F ± 2°F)<br />

40°C = 43° ± 1°C (109°F ± 2°F)<br />

44°C = 47° ± 1°C (117°F ± 2°F)<br />

Electrical<br />

Supply Power Input:<br />

Type<br />

100V 100 VAC, 50/60 Hz, 9.00 Amps<br />

115V 115 VAC, 50/60 Hz, 8.05 Amps<br />

230V 230 VAC, 50/60 Hz, 4.0 Amps<br />

Protection Against Electrical Shock Class 1 Equipment<br />

Type BF<br />

EQUATOR ® <strong>Convective</strong> <strong>Warmer</strong> | Operator’s Manual 33