S E C T I O N 8 • M a i n t e n a n c e<br />
Test 5: Low (36°C) Temperature Test<br />
1 Press the Low button , and allow the temperature to stabilize for at least 5 minutes.<br />
2 After the temperature has stabilized, verify that the Low temperature readings appear as 36°C.<br />
Test 6: Low (36°C) Over Temperature and Under Temperature Test<br />
1 Press the OFF button and then press the ON button . The device will restart and proceed<br />
through the self-test sequence.<br />
2 If the device is hot, let it cool to approximately 30°C before proceeding.<br />
3 Press and hold the LOW button on the front panel for approximately 8 seconds, until the Low<br />
button LED starts to flash.<br />
4 As the temperature increases, verify that the following events occur:<br />
• Between 32°C and 34°C, the Under Temperature LED ( ) stops flashing.<br />
• Between 38°C and 40°C, the Over Temperature alarm sounds, the Over Temperature LED ( )<br />
lights, the motor stops, and the control panel displays the temperature at which the Over<br />
Temperature alarm was activated.<br />
Test 7: Medium (40°C) Over Temperature and Under Temperature Test<br />
1 Press the OFF button and then press the ON button . The device will restart and proceed<br />
through the self-test sequence.<br />
2 If the device is hot, let it cool to approximately 35°C before proceeding.<br />
3 Press and hold the Medium button on the control panel for approximately 8 seconds, until the<br />
Medium button LED starts to flash.<br />
4 As the temperature increases, verify that the following events occur:<br />
• Between 36°C and 38°C, the Under Temperature LED ( ) stops flashing.<br />
• Between 42°C and 44°C, the Over Temperature alarm sounds, the Over Temperature LED ( )<br />
lights, the motor stops, and the control panel displays the temperature at which the Over<br />
Temperature alarm was activated.<br />
Test 8: High (44°C) Over Temperature and Under Temperature Test<br />
1 Press the OFF button and then press the ON button . The device will restart and proceed<br />
through the self-test sequence.<br />
2 If the device is hot, let it cool to approximately 40°C before proceeding.<br />
26 EQUATOR ® <strong>Convective</strong> <strong>Warmer</strong> | Operator’s Manual