Controls, Indicators, and Displays<br />

S E C T I O N 5 • O p e r a t i o n<br />

The following table defines each control, indicator, and display and the modes of operation.<br />

Control/Indicator Definition<br />

Power ON Button<br />

The green button powers ON the convective warmer.<br />

Power OFF Button<br />

The orange button powers OFF operating components of the convective warmer.<br />

Note: To remove all power from the convective warmer, the MAINS power cord must be<br />

removed from the electrical receptacle.<br />

Ambient Air Temperature Setting Button<br />

When selected, the heater is turned off and the motor still operates (air flows).<br />

36°C Temperature Setting Button<br />

When selected, the air temperature is set to 36°C at the hose end.<br />

40°C Temperature Setting Button<br />

When selected, the air temperature is set to 40°C at the hose end.<br />

44°C Temperature Setting Button<br />

When selected, the air temperature is set to 44°C at the hose end.<br />

Over Temperature Alarm Indicator<br />

If the hose end temperature is more than 3°C (± 1°C) above the selected temperature, the<br />

indicator illuminates, a continuous audible alarm sounds, and the heater and motor shut off.<br />

Under Temperature Indicator<br />

If the hose end temperature is more than 3°C (± 1°C) below the selected temperature, this<br />

indicator flashes.<br />

Disconnect Indicator<br />

If the thermistor cable or the hose is not attached properly, this indicator illuminates, an<br />

audible alarm sounds and repeats approximately every five seconds, and the heater and<br />

motor shut off.<br />

If the heater has an open wire circuit, this indicator illuminates, an audible alarm sounds and<br />

repeats approximately every five seconds, the heater shuts off, but the motor continues to<br />

operate (air flows).<br />

Hose End Temperature Display<br />

Displays the temperature of the air at the hose end.<br />