Controls, Indicators, and Displays<br />
S E C T I O N 5 • O p e r a t i o n<br />
The following table defines each control, indicator, and display and the modes of operation.<br />
Control/Indicator Definition<br />
Power ON Button<br />
The green button powers ON the convective warmer.<br />
Power OFF Button<br />
The orange button powers OFF operating components of the convective warmer.<br />
Note: To remove all power from the convective warmer, the MAINS power cord must be<br />
removed from the electrical receptacle.<br />
Ambient Air Temperature Setting Button<br />
When selected, the heater is turned off and the motor still operates (air flows).<br />
36°C Temperature Setting Button<br />
When selected, the air temperature is set to 36°C at the hose end.<br />
40°C Temperature Setting Button<br />
When selected, the air temperature is set to 40°C at the hose end.<br />
44°C Temperature Setting Button<br />
When selected, the air temperature is set to 44°C at the hose end.<br />
Over Temperature Alarm Indicator<br />
If the hose end temperature is more than 3°C (± 1°C) above the selected temperature, the<br />
indicator illuminates, a continuous audible alarm sounds, and the heater and motor shut off.<br />
Under Temperature Indicator<br />
If the hose end temperature is more than 3°C (± 1°C) below the selected temperature, this<br />
indicator flashes.<br />
Disconnect Indicator<br />
If the thermistor cable or the hose is not attached properly, this indicator illuminates, an<br />
audible alarm sounds and repeats approximately every five seconds, and the heater and<br />
motor shut off.<br />
If the heater has an open wire circuit, this indicator illuminates, an audible alarm sounds and<br />
repeats approximately every five seconds, the heater shuts off, but the motor continues to<br />
operate (air flows).<br />
Hose End Temperature Display<br />
Displays the temperature of the air at the hose end.<br />
EQUATOR ® <strong>Convective</strong> <strong>Warmer</strong> | Operator’s Manual 13