S E C T I O N 8 • M a i n t e n a n c e<br />
3 Press and hold the High button on the control panel for approximately 8 seconds, until the High<br />
button LED starts to flash.<br />
4 As the temperature increases verify that the following events occur:<br />
• Between 40°C and 42°C, the Under Temperature LED ( ) stops flashing.<br />
• Between 46°C and 48°C, the Over Temperature alarm sounds, the Over Temperature LED ( )<br />
lights, the motor stops, and the control panel displays the temperature at which the Over<br />
Temperature alarm was activated.<br />
Test 9: Ambient Over Temperature Test<br />
1 Press the OFF button and then press the ON button . The device will restart and proceed<br />
through the self-test sequence.<br />
2 If the device is hot, let it cool to approximately 44°C before proceeding.<br />
3 Press and hold the Ambient button on the control panel until the Ambient button LED starts to<br />
flash.<br />
4 As the temperature increases, verify that the following events occur:<br />
• Between 46°C and 48°C, the Over Temperature alarm sounds, the Over Temperature LED ( )<br />
lights, the motor stops, and the control panel displays the temperature at which the Over<br />
Temperature alarm was activated.<br />
5 Press the OFF button .<br />
EQUATOR ® Convective Warmer | Operator’s <strong>Manual</strong> 27