S E C T I O N 8 • M a i n t e n a n c e<br />
3 Check that all the LEDs turn off, except the Ambient button LED, and that the motor continues to run.<br />
4 The self-test is completed. If the self-test sequence did not complete properly, refer to the EQUATOR ®<br />
Convective Warmer Service <strong>Manual</strong>.<br />
Test 2: Disconnect Indicator Test<br />
1 If necessary, press the OFF button , then press the ON button . Wait for the self-test to complete.<br />
2 Disconnect the hose thermistor cable from the rear of the device. Verify that the Disconnect LED ( )<br />
lights, the alarm beeps, and the motor stops.<br />
3 Press the OFF button and then press the ON button . The self-test should stop on the first test.<br />
4 Reconnect the thermistor cable. The self-test should continue to completion.<br />
5 Disconnect the hose from the rear of the device. Verify that the Disconnect LED ( ) lights, the<br />
alarm beeps, and the motor stops.<br />
6 Reconnect the hose. Verify that the Disconnect LED ( ) turns off and the device continues to run.<br />
7 Press the High button . Wait for the temperature to rise to approximately 30°C.<br />
8 Disconnect the hose from the rear of the device. Verify that the Disconnect LED ( ) lights, the<br />
alarm beeps, and the motor stops.<br />
9 Reconnect the hose. Verify that the Disconnect LED ( ) turns off, the motor restarts, and the<br />
temperature continues to increase toward 44°C.<br />
10 Press the Ambient button .<br />
Test 3: High (44°C) Temperature Test<br />
This test verifies the High temperature readings.<br />
1 Press the High button and allow the temperature to stabilize for at least 5 minute.<br />
2 After the temperature has stabilized, verify that the High temperature readings appear as 44°C.<br />
Test 4: Medium (40°C) Temperature Test<br />
1 Press the Medium button , and allow the temperature to stabilize for at least 5 minutes.<br />
2 After the temperature has stabilized, verify that the Medium temperature readings appear as 40°C.<br />
EQUATOR ® Convective Warmer | Operator’s <strong>Manual</strong> 25