7 After thoroughly cleaning the convective warmer, perform disinfection if required.<br />
S E C T I O N 8 • M a i n t e n a n c e<br />
8 If it is hospital policy to perform disinfection as part of reprocessing, then follow your institution's<br />
guidelines for disinfecting of the surfaces of non-critical medical devices. The list below includes lowlevel<br />
disinfectants that are commonly used in the medical community and high-level disinfectants that<br />
are claimed by the manufacturer. The effectiveness of these listed disinfectants should be validated using<br />
the hospital procedures.<br />
The following disinfectant agents can be used without causing damage to the enclosure:<br />
Low Level Disinfectants:<br />
• fantastik ® All Purpose Cleaner<br />
High Level Disinfectants:<br />
• 1.56% Phenol (e.g., Sporicidin ® )<br />
• 3.4% Glutaraldehyde (e.g., CIDEX ® Plus)<br />
• 10% Bleach solution<br />
• 1% Ammonia solution<br />
• Surface disinfectants compatible with plastic or metal materials.<br />
9 Rinsing of the disinfectant residue should be done using a soft cloth or sponge as the applicator.<br />
Replace the Air Filter (F3-5000)<br />
Replace the air filter every 12 months, or sooner if required.<br />
Note: Always perform functional tests after replacing the air filter.<br />
WARNINGS<br />
• Do not change the air filter while the convective warmer is operating. Risk of thermal injury<br />
exists if the device is operating.<br />
EQUATOR ® Convective Warmer | Operator’s <strong>Manual</strong> 23