S E C T I O N 4 • A s s e m b l y I n s t r u c t i o n s<br />
WARNINGS<br />
• Do not mount the convective warmer higher than 117cm (46") on the IV pole. For convenience,<br />
117cm (46") is indicated by a black mark on the power cord. Mounting the convective warmer<br />
above 117cm (46") may result in instability of the pole and cause tipping that may injure the<br />
patient or user.<br />
• Ensure that the convective warmer I.V. pole mounting clamp is properly tightened before each<br />
use. Failure to securely mount the convective warmer onto the I.V. pole may cause it to slide<br />
down the I.V. pole, and injure the patient or user.<br />
Mount to an I.V. Pole<br />
Note: Do not place the convective warmer higher than 117cm (46")<br />
from the floor.<br />
1 Place the convective warmer, on the pole and turn the clamp (d) to<br />
tighten.<br />
2 Check the tightness of the Convective Warmer to ensure it is securely<br />
clamped to the pole.<br />
WARNINGS<br />
• Grounding reliability can only be achieved when the MAINS power cord is connected to a<br />
properly grounded receptacle. Risk of electrical shock exists if the equipment is not connected<br />
to a properly grounded receptacle resulting in death or serious injury to the patient or user.<br />
Install the Power Cord<br />
1 Insert the plug on the power cord (e) into the power cord receptacle on<br />
the rear of the convective warmer.<br />
2 Push the power cord lock over the plug to secure it in the receptacle.<br />
Step 4: Perform Electrical Safety Tests<br />
Perform all applicable electrical safety tests as required per institutional procedure. Electrical safety check<br />
must be performed by competent personnel authorized by the institution to perform such testing. These tests<br />
include, but are not limited to: leakage current, ground bond test, and hypot. Refer to the EQUATOR <br />
Convective Warmer Service <strong>Manual</strong> (PN 4533902-GB) for further information about electrical safety testing.<br />
Perform electrical safety checks on a routine basis according to institutional policy.<br />