S E C T I O N 4 • A s s e m b l y I n s t r u c t i o n s<br />

WARNINGS<br />

• Do not mount the convective warmer higher than 117cm (46") on the IV pole. For convenience,<br />

117cm (46") is indicated by a black mark on the power cord. Mounting the convective warmer<br />

above 117cm (46") may result in instability of the pole and cause tipping that may injure the<br />

patient or user.<br />

• Ensure that the convective warmer I.V. pole mounting clamp is properly tightened before each<br />

use. Failure to securely mount the convective warmer onto the I.V. pole may cause it to slide<br />

down the I.V. pole, and injure the patient or user.<br />

Mount to an I.V. Pole<br />

Note: Do not place the convective warmer higher than 117cm (46")<br />

from the floor.<br />

1 Place the convective warmer, on the pole and turn the clamp (d) to<br />

tighten.<br />

2 Check the tightness of the Convective Warmer to ensure it is securely<br />

clamped to the pole.<br />

WARNINGS<br />

• Grounding reliability can only be achieved when the MAINS power cord is connected to a<br />

properly grounded receptacle. Risk of electrical shock exists if the equipment is not connected<br />

to a properly grounded receptacle resulting in death or serious injury to the patient or user.<br />

Install the Power Cord<br />

1 Insert the plug on the power cord (e) into the power cord receptacle on<br />

the rear of the convective warmer.<br />

2 Push the power cord lock over the plug to secure it in the receptacle.<br />

Step 4: Perform Electrical Safety Tests<br />

Perform all applicable electrical safety tests as required per institutional procedure. Electrical safety check<br />

must be performed by competent personnel authorized by the institution to perform such testing. These tests<br />

include, but are not limited to: leakage current, ground bond test, and hypot. Refer to the EQUATOR <br />

Convective Warmer Service <strong>Manual</strong> (PN 4533902-GB) for further information about electrical safety testing.<br />

Perform electrical safety checks on a routine basis according to institutional policy.<br />

EQUATOR ® Convective Warmer | Operator’s <strong>Manual</strong> 11<br />