Operator's Manual - Smiths Medical
Operator's Manual - Smiths Medical
Operator's Manual - Smiths Medical
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
S E C T I O N 3 • I m p o r t a n t S a f e t y I n f o r m a t i o n<br />
WARNINGS (continued)<br />
• Do not sterilize convective warming blankets. Sterilizing blankets may damage the blanket<br />
and may cause thermal injury to the patient.<br />
• If Over Temperature audible alarm sounds and/or red Over Temperature alarm indicator<br />
illuminates, discontinue use of the convective warmer to prevent thermal injury to the patient.<br />
Remove the device from service. Contact <strong>Smiths</strong> <strong>Medical</strong> or your local <strong>Smiths</strong> <strong>Medical</strong><br />
distributor.<br />
• If the convective warmer does not perform its self-test properly, fails to operate, or stops<br />
while running, discontinue use of the convective warmer to prevent thermal injury to the<br />
patient. Remove the device from service. Contact <strong>Smiths</strong> <strong>Medical</strong> or your local <strong>Smiths</strong> <strong>Medical</strong><br />
distributor.<br />
• The convective warmer must be calibrated by competent personnel authorized by <strong>Smiths</strong><br />
<strong>Medical</strong>. Failure to calibrate the device correctly may result in thermal injury to the patient.<br />
• Do not use the SW-2008 convective warming blanket if the package is damaged or open.<br />
Sterility may be compromised.<br />
CAUTIONS<br />
Malfunction, failure, or damage to the device may occur if these cautions are not followed:<br />
• Do not use alcohol or other strong solvents to clean the hose or exterior surfaces. These<br />
solutions may damage labels and other parts.<br />
• Never use organic solvents (e.g., acetone), strong acids, or bases to clean any portion of the<br />
convective warmer.<br />
• Do not place the convective warmer directly under a faucet or use a faucet sprayer to rinse.<br />
Never spray cleaning or other fluids into openings on the convective warmer or into the<br />
external connectors.<br />
• <strong>Medical</strong> devices require specific material characteristics to perform as intended. These<br />
characteristics have been verified for single use only. Any attempt to re-process the device for<br />
subsequent re-use may adversely affect the integrity of the device or lead to deterioration in<br />
performance.<br />
• Federal (U.S.A.) law restricts this device to sale by or on the order of a physician.<br />
8 EQUATOR ® Convective Warmer | Operator’s <strong>Manual</strong>