<strong>6700</strong> SAMPLERS<br />

Chapter 5 Reference<br />

Chapter Contents<br />

Sample Event<br />

Sample Volume<br />

SuctionHead<br />

Rinses and Retries<br />

Pacing<br />

Distribution<br />

<strong>Sampler</strong> Enable<br />

Pauses/Resumes<br />

StartTimes<br />

Other Functions<br />

<strong>Manual</strong> Functions<br />

ProgrammingStyle<br />

Hardware Setup<br />

SystemIDs<br />

Memory<br />

Pressurized Lines<br />

Command Driven Operation<br />

1<br />

2<br />

Extended Screen for<br />

Suction Head Settings<br />

AUTO SUCTION HEAD<br />

ENTER HEAD<br />

SUCTION HEAD OF<br />

___ ft<br />

(min - max)<br />

81<br />

SAMPLE EVENT<br />

A sample event consists of a complete sampling cycle. A sample<br />

event includes presample and postsample purges, line rinses,<br />

sampling retries, and deliveries of sample volumes. Only one<br />

sample volume is placed in a given bottle during a sample event,<br />

however, depending on the distribution, multiple bottles may<br />

receive sample volumes.<br />

SAMPLE VOLUME<br />

The sample volume is the discrete, programmed amount of liquid<br />

delivered. During a sample event the sampler may deliver a sample<br />

volume to several bottles, depending on the distribution.<br />

“Sample volumes” and “samples,” are used interchangeably.<br />

Extended programming adds a flow-dependent sample volume<br />

option for uniform time-paced programs. If a flow module is<br />

attached, the flow source may be the module’s flow volume. Otherwise<br />

it is the flow pulse count received at the external flow<br />

meter connector. You enter the amount of flow required for each<br />

10 ml of liquid. When the sample event is initiated, the sample<br />

volume will be based on the flow that has occurred since the last<br />

sample. This sample volume is will be at least 20 ml, but not<br />

more than the bottle volume (or 9990 ml, whichever is smaller).<br />

With flow-dependent sample volumes, the sampler resets the<br />

flow volume count at the start time. Therefore, the sampler will<br />

not take a sample at the start time.<br />

SUCTION HEAD<br />

Suction head is the vertical distance between the surface of the<br />

liquid source and the sampler’s pump. At each sample event, the<br />

sampler determines the suction head automatically.<br />

In extended programming, you can manually enter the suction<br />

head if you want. Select ENTER HEAD only when the head<br />

remains stable and you can measure it accurately.