<strong>6700</strong> SAMPLERS<br />

Example 8 Resetting the Internal Battery Warning<br />

1<br />

2<br />

3<br />

4<br />

5<br />

6<br />

The circuit boards can be<br />

damaged from a discharge<br />

of static electricity. Always<br />

use an antistatic mat when<br />

opening the controller to<br />

protect the circuit boards.<br />

Note<br />

CAUTION<br />

Poor soldering techniques can<br />

easily damage a printed circuit<br />

board. The battery should be<br />

replaced by a skilled technician.<br />

RUN<br />

PROGRAM<br />

VIEW REPORT<br />

OTHER FUNCTIONS<br />

MAINTENANCE<br />

MANUAL FUNCTIONS<br />

PROGRAMMING STYLE<br />

HARDWARE<br />

SET CLOCK<br />

SET PUMP ALARM<br />

INTERNAL BATTERY<br />

DIAGNOSTICS<br />

WARNING!<br />

REPLACE INTERNAL<br />

BATTERY BEFORE<br />

17-OCT-97<br />

DID YOU REPLACE THE<br />

INTERNAL BATTERY?<br />

YES NO<br />

MAINTENANCE<br />

MANUAL FUNCTIONS<br />

PROGRAMMING STYLE<br />

HARDWARE<br />

3. Remove the EMI shield and grounding straps by loosening the<br />

five 1 /4-inch stop nuts.<br />

4. Disconnect the module, liquid detector, and pump count connectors<br />

from the main circuit board (Figure 20).<br />

5. Remove the three screws and the 1 /4-inch stud attaching the<br />

main circuit board to the case.<br />

6. Detach the main circuit board from the keypad connector by<br />

pulling gently upward on the edges of the board next to the<br />

connector. Pull the main board up until the keypad connector<br />

is almost even with the edge of the case. Next, detach the main<br />

circuit board from the connector board.<br />

7. Desolder the lithium battery, using care to prevent damage to<br />

the circuit board. Insert the replacement battery and solder it<br />

in place.<br />

8. Reconnect the main circuit board to the connector board before<br />

reconnecting it to the keypad. Replace the 1 /4-inch stud and<br />

the three screws, and reconnect the module, liquid detector,<br />

and pump count connectors.<br />

9. Replace the EMI shield and grounding straps. Reconnect the<br />

two distributor connectors.<br />

10. Ensure that all connections and hardware are secure.<br />

Replace the desiccant and close the case.<br />

11. Reset the battery alarm.<br />

Select OTHER FUNCTIONS.<br />

Select MAINTENANCE.<br />

Select INTERNAL BATTERY.<br />

The sampler displays this screen briefly to show you the expiration date.<br />

After replacing the battery, select YES. Select NO only if checking the<br />

expiration date.<br />

Resetting the alarm directs the sampler to calculate a new expiration date,<br />

five years in the future. If you reset the alarm without replacing the battery,<br />

the sampler will not be able to alert you to the battery’s impending<br />

expiration.<br />

Press Stop to return to the main menu.<br />