<strong>6700</strong> SAMPLERS<br />
Example 8 Resetting the Internal Battery Warning<br />
1<br />
2<br />
3<br />
4<br />
5<br />
6<br />
The circuit boards can be<br />
damaged from a discharge<br />
of static electricity. Always<br />
use an antistatic mat when<br />
opening the controller to<br />
protect the circuit boards.<br />
Note<br />
CAUTION<br />
Poor soldering techniques can<br />
easily damage a printed circuit<br />
board. The battery should be<br />
replaced by a skilled technician.<br />
RUN<br />
PROGRAM<br />
VIEW REPORT<br />
OTHER FUNCTIONS<br />
MAINTENANCE<br />
MANUAL FUNCTIONS<br />
PROGRAMMING STYLE<br />
HARDWARE<br />
SET CLOCK<br />
SET PUMP ALARM<br />
INTERNAL BATTERY<br />
DIAGNOSTICS<br />
WARNING!<br />
REPLACE INTERNAL<br />
BATTERY BEFORE<br />
17-OCT-97<br />
DID YOU REPLACE THE<br />
INTERNAL BATTERY?<br />
YES NO<br />
MAINTENANCE<br />
MANUAL FUNCTIONS<br />
PROGRAMMING STYLE<br />
HARDWARE<br />
3. Remove the EMI shield and grounding straps by loosening the<br />
five 1 /4-inch stop nuts.<br />
4. Disconnect the module, liquid detector, and pump count connectors<br />
from the main circuit board (Figure 20).<br />
5. Remove the three screws and the 1 /4-inch stud attaching the<br />
main circuit board to the case.<br />
6. Detach the main circuit board from the keypad connector by<br />
pulling gently upward on the edges of the board next to the<br />
connector. Pull the main board up until the keypad connector<br />
is almost even with the edge of the case. Next, detach the main<br />
circuit board from the connector board.<br />
7. Desolder the lithium battery, using care to prevent damage to<br />
the circuit board. Insert the replacement battery and solder it<br />
in place.<br />
8. Reconnect the main circuit board to the connector board before<br />
reconnecting it to the keypad. Replace the 1 /4-inch stud and<br />
the three screws, and reconnect the module, liquid detector,<br />
and pump count connectors.<br />
9. Replace the EMI shield and grounding straps. Reconnect the<br />
two distributor connectors.<br />
10. Ensure that all connections and hardware are secure.<br />
Replace the desiccant and close the case.<br />
11. Reset the battery alarm.<br />
Select OTHER FUNCTIONS.<br />
Select MAINTENANCE.<br />
Select INTERNAL BATTERY.<br />
The sampler displays this screen briefly to show you the expiration date.<br />
After replacing the battery, select YES. Select NO only if checking the<br />
expiration date.<br />
Resetting the alarm directs the sampler to calculate a new expiration date,<br />
five years in the future. If you reset the alarm without replacing the battery,<br />
the sampler will not be able to alert you to the battery’s impending<br />
expiration.<br />
Press Stop to return to the main menu.<br />
80