CAUTION<br />

Desiccant may produce<br />

irritating fumes when heated.<br />

Observe the following<br />

precautions:<br />

Useaventedoveninawell<br />

ventilated room.<br />

Do not remain in the room<br />

while recharging is taking<br />

place.<br />

Use the recommended<br />

temperature.<br />

There have been reports of<br />

irritating fumes coming from<br />

the desiccant during reactivation.<br />

While our attempts to<br />

duplicate the problem have<br />

been unsuccessful, we still<br />

urge you to use caution.<br />

Material Safety Data Sheets<br />

are provided in Appendix B.<br />

Battery Warning<br />

WARNING!<br />

REPLACE INTERNAL<br />

BATTERY BEFORE<br />

17-OCT-97<br />

Note<br />

<strong>Isco</strong> recommends you replace<br />

the battery every five years, as<br />

soon as the battery warning<br />

appears. To insure that the<br />

replacement meets <strong>Isco</strong>’s<br />

specifications, use only <strong>Isco</strong><br />

replacement batteries. (See<br />

Appendix D, Replacement<br />

Parts.)<br />

<strong>6700</strong> SAMPLERS<br />

RECHARGING THE DESICCANT<br />

The <strong>6700</strong> <strong>Sampler</strong>s use a desiccant bag inside the controller case<br />

to prevent moisture damage to its electronic components.<br />

Follow the instructions in Opening the Controller Case on page<br />

77 to open the controller. Remove the bag from the cardboard box<br />

atthefrontofthecase(Figure 19).<br />

How To Recharge The Desiccant<br />

Do not put the cardboard box in the oven. You should recharge<br />

the desiccant when the area marked “30” on the paper humidity<br />

indicator turns pink.<br />

Place a sheet of brown paper on a flat metal sheet. You can use a<br />

brown grocery bag and a typical cookie sheet. Place only the bags<br />

on the sheet. Do not stack the bags on top of each other or allow<br />

them to touch. Place in a vented, circulating forced air, convection<br />

oven in a well ventilated room. Allow two inches of air space<br />

between the top of the bags and the next metal tray above the<br />

bags. Keep the tray a minimum of 16 inches from heating element.<br />

Heat the bags at a temperature of 240 to 250oF(116to 121oC) for 12 to 16 hours. At the end of the time period, the bags<br />

should be immediately removed and placed in an air tight container<br />

for cooling. The desiccant will be recharged to approximately<br />

80 to 90% of its previous capacity. After repeated<br />

recharging, the desiccant bag may require replacement.<br />

Some bags will have the temperature and time for recharging the<br />

desiccant printed on the bag. If they differ, use the temperature<br />

and time printed on the bag.<br />

REPLACING THE INTERNAL BATTERY<br />

A lithium battery housed inside the controller maintains power<br />

to the sampler’s memory when the controller is disconnected<br />

from a power source. If it discharges completely, the sampler will<br />

lose all program settings and all data stored in memory when the<br />

external power source is disconnected. The battery requires<br />

replacement every five years. The sampler displays a warning<br />

withinonemonthofitsexpectedlifetoalertyoutothebattery’s<br />

impending expiration. The warning appears each time you turn<br />

the sampler on until you reset the alarm. When you reset the<br />

alarm, the sampler resets the warning to a date 59 months from<br />

the current date.<br />

While replacing the battery, it is likely that the stored readings<br />

will be lost and your program will revert to the default program.<br />

Before proceeding, retrieve the stored data (see Collecting<br />

Reports on page 59) and record your program settings.<br />

To replace the internal battery:<br />

1. Open the controller case (see Opening the Controller Case on<br />

page 77). Separate the case bottom and top by removing the<br />

two distributor connectors (Figure 19).<br />

2. Lift the desiccant box from the case (Figure 19).<br />

79