<strong>6700</strong> SAMPLERS<br />

Example 7 Resetting the Pump Tube Warning<br />

1<br />

2<br />

3<br />

4<br />

5<br />

6<br />

7<br />

RUN<br />

PROGRAM<br />

VIEW REPORT<br />

OTHER FUNCTIONS<br />

MAINTENANCE<br />

MANUAL FUNCTIONS<br />

PROGRAMMING STYLE<br />

HARDWARE<br />

SET CLOCK<br />

SET PUMP ALARM<br />

INTERNAL BATTERY<br />

DIAGNOSTICS<br />

1024000 PUMP COUNTS,<br />

WARNING AT 1000000<br />

1024000 PUMP COUNTS,<br />

RESET PUMP COUNTER?<br />

YES NO<br />

WARNING AT 1000000<br />

PUMP COUNTS<br />

(1 - 99)00000)<br />

MAINTENANCE<br />

MANUAL FUNCTIONS<br />

PROGRAMMING STYLE<br />

HARDWARE<br />

The circuit boards can be<br />

damaged from a discharge<br />

of static electricity. Always<br />

use an antistatic mat when<br />

opening the controller to<br />

protect the circuit boards.<br />

Note<br />

CAUTION<br />

The case top and bottom fit<br />

together with a sealed tongueand-groove<br />

joint. Any damage<br />

— nicks or cuts — to the<br />

tongue, groove, or sealing<br />

gasket prevent the case from<br />

sealing completely when you<br />

close it. Use extreme care when<br />

opening the case to avoid<br />

damaging the joining surfaces.<br />

Select OTHER FUNCTIONS.<br />

Select MAINTENANCE.<br />

Select SET PUMP ALARM.<br />

The sampler displays this screen briefly to give you the current pump-count<br />

information. The first line shows you the pump counts tallied since the<br />

counter was last reset. The second line shows the current alarm setting.<br />

To reset the counter to zero, select YES. Always reset the counter after<br />

replacing a pump tube. Select NO when merely checking the current count.<br />

If necessary, change the pump-count alarm setting by typing the first two<br />

digits of the new setting. The sampler accepts entries between 1 and 99. For<br />

example, to increase the count to 1,500,000, enter 15.<br />

Press Stop to return to the main menu.<br />

OPENING THE CONTROLLER CASE<br />

You will need to open the controller case to gain access to:<br />

The desiccant bag that dries the controller’s interior<br />

The battery that provides backup power to the controller<br />

To open the case:<br />

1. Pull the discharge tube and pump tube away from the bulkhead<br />

fitting.<br />

2. Remove the distributor arm by unscrewing the nut that<br />

attaches the arm to the distributor shaft. DO NOT ROTATE<br />

THE DISTRIBUTOR ARM MANUALLY; THIS WILL<br />

SEVERELY DAMAGE THE DISTRIBUTOR DRIVE<br />

GEARS.<br />

3. Remove the controller from the sampler.<br />

4. Open the case by removing the 11 screws that attach the case<br />

bottom to the case top.<br />

Refer to Figure 19 for the location of the desiccant box. To gain<br />

access to the main circuit board and battery, remove the two distributor<br />

connectors and the EMI shield.<br />

77