Note<br />

For critical sampling applications,<br />

consider replacing the<br />

suction line, pump tube, and<br />

discharge tube. Replacement<br />

eliminates the possibility of any<br />

cross contamination from previous<br />

sampling sites.<br />

<strong>6700</strong> SAMPLERS<br />

CLEANING PROTOCOLS FOR PRIORITY<br />

POLLUTANTS<br />

Clean sampling equipment is essential for valid laboratory analysis.<br />

<strong>Isco</strong> recommends that you develop cleaning protocols in consultation<br />

with a laboratory analyst when designing the<br />

monitoring program. For example, to clean sample bottles, suction<br />

line, and pump tubes, Lair (1974) suggests these protocols used by<br />

USEPA Region IV Surveillance and Analysis field personnel<br />

engaged in NPDES compliance monitoring. The protocols are<br />

based on U.S. Environmental Protection Agency Publications<br />

EPA-600/4-77-039 (Sampling of Water and Wastewater by Dr.<br />

Phillip E. Shelley).<br />

<strong>Isco</strong> Glass Sample Bottles<br />

1. One spectro-grade acetone rinse.<br />

2. Dishwasher cycle (wash and tap water rinse, no detergent).<br />

3. Acid wash with at least 20 percent hydrochloric acid.<br />

4. Dishwasher cycle (wash and tap water rinse, no detergent).<br />

5. Replace in covered <strong>Isco</strong> tubs.<br />

Vinyl Suction Line<br />

1. Use new suction line for each new sampling setup, therefore,<br />

no cleaning is required.<br />

2. Use Teflon suction line when sampling for organic compounds.<br />

Teflon Suction Line<br />

1. Rinse twice with spectro-grade acetone.<br />

2. Rinse thoroughly with hot tap water using a brush, if possible,<br />

to remove particulate matter and surface film.<br />

3. Rinse thoroughly three times with tap water.<br />

4. Acid wash with at least 20 percent hydrochloric acid.<br />

5. Rinse thoroughly three times with tap water.<br />

6. Rinse thoroughly three times with distilled water.<br />

7. Rinse thoroughly with petroleum ether and dry by pulling air<br />

through the line.<br />

8. Dry overnight in a warm oven (use an oven temperature of<br />

lower than 150 degrees F), if possible.<br />

9. Cap ends with aluminum foil.<br />

<strong>Isco</strong> Pump Tube<br />

1. Rinse by pumping hot tap water through the tube for at least 2<br />

minutes.<br />

2. Acid wash the tube by pumping at least a 20 percent solution<br />

of hydrochloric acid through the tube for at least 2 minutes.<br />

3. Rinse by pumping hot tap water through the tube for at least 2<br />

minutes.<br />

4. Rinse by pumping distilled water through the tube for at least<br />

2minutes.<br />